DRAWING NO:     1080023 
REVISION:            34 

 

CLASSIFICATION: 

 

1. Protection against electric shock – Class I. 

 

2. Protection against harmful ingress of water – IPXO (Non-protected), ordinary. 

 

3. Methods of sterilization – None. 
4. Suitability for use in an oxygen rish environment – End user responsibility, not evaluated. 
5. Mode of operation – Continuous. 

 

WARNING! 

An open internal fuse indicates a catastrophic failure of circuit component(s). Repair must be by authorized IPD personnel only.   

RISK OF FIRE! 

Refer to fuse rating on power supply circuit board for rating.  

 

WARNING! 

Dangerous voltages are present on some components, printed circuit board traces and heatsinks. 

SHOCK HAZARD! 

 

INPUT FUSE: 

This product includes a single fuse in the phase lead only. In consideration of IEC 60601-1:2005 Clause 8.11.5, a second fuse may 
be required in the neutral lead of the end use equipment 

 

SEPERATION: 

Primary  to  secondary  creepage  distance  is  8mm  minimum,  clearance  5mm  minimum  (2MOPP).  Primary  to  ground  creepage 
distance is 4mm minimum, clearance 2.5mm minimum (1MOPP). Secondary to ground creepage is 2.3mm minimum, clearance is 
1.4mm  minimum  (Operational  Insulation).  Required  seperation  must  be  maintained  in  the  end  use  equipment  to  preserve  the 
established means of protection.  

 

OUTPUTS:

 

The  outputs  are  not  acceptable  for  patient  connection  without  additional  isolation.  All  outputs  are  ES1  under  normal  and  single 
fault conditions unless otherwise indicated. 

 

TEMPERATURES:  

The  maximum operating  temperatures  of  safety  components  as  defined  in  the  applicable  safety  standards  must  not  be  exceeded 
after installation in the end use equipment. Output power, ambient air temperature and convection or forced air cooling availability 
should be considered in the end use equipment. 

 

HIPOT: 

In consideration of IEC 60601-1:2005 Clause 8.8.3, care must be taken to insure the voltage applied to a reinforced insulation does 
not overstress different types and levels of insulation. Breakdown of basic insulation and catastrophic failure of the power supply 
may result if a test voltage of greater than 1800 VAC is applied between primary and secondary circuits. Each isolating component 
is  factory  tested  at  4000  VAC  minimum  prior  to  installation.  Primary  and  secondary  to  ground  capacitors  may  need  to  be 
disconnected prior to performing a dielectric strength test on the power supply or the end product. 

 

 

INSTALLATION: 

The power supplies included in model listing on reverse side are considered components intended for professional installation into 
end use equipment. The protective earth (ground) terminal must be bonded to protective earth in the end use equipment.  

 

EMISSIONS:

 

This  product  was  tested  for  compliance  with  EN  55022  and  EN  55011  Class  B  conducted  and  radiated  emissions  using  the 
techniques listed below and non-inductive load resistors to simulate operation in a typical installation. All or a combination of the 
following requirements may be necessary to insure compliance in the end use equipment.  

 

 

 

 

 

 

1. Installation of the power supply, output cables and loads in a shielded enclosure. 

   

2. Use of optional chassis and cover. 

   

3. Use of shielded I/O cables. 

   

4. Use of ferrite beads on I/O cables. 

   

5. Grounded output returns. 

 
CONNECTIONS /

 

 

DIMENSIONS: 

   

 

 

 

 

 

 
 
 
 
 
 
 
 

 

 
AIR FLOW      

1. Recommended     2. Good     3. Fair

 

DIRECTION: 

 

CONNECTORS: 

P1:  AC Input – 0.156 inch friction lock header mates with Molex 09-50-3031 or equivalent crimp terminal housing with Molex 

2478 or equivalent crimp terminal. 

P2:  DC Output (Single Output) – 6-32 screw-down terminal mates with #6 ring-tongue terminal (10 in.-lb. max.). 
P2:  DC Output (Multiple Output)  – 0.156 inch friction lock header mates with Molex 09-50-3161 or equivalent crimp terminal 

housing with Molex 2478 or equivalent crimp terminal. 

G: 

Protective Earth (Ground) – 0.187 quick disconnect terminal. 

P3:  Option  /  Sense  (Single  Output)  –  0.100  friction  lock  header  mates  with  Molex  50-57-9008  or  equivalent  crimp  terminal 

housing with Molex type 71851 or equivalent crimp terminal. 

P3:  Option  /  Sense  (Multiple  Output)  -  0.100  breakaway  header  mates  with  Molex  22-55-2081  or  equivalent  crimp  terminal 

housing with Molex type 71851 or equivalent crimp terminal. 