Quick Start Guide
VIAFLO/VOYAGER Electronic Pipettes
This quick start guide is intended to provide a quick overview of your pipette’s key
features and to offer basic instructions for getting started. For detailed information,
please refer to the latest operating instructions that can be found at
Intended use
This is a general-purpose laboratory instrument for use in research only.
VIAFLO/VOYAGER
Electronic Pipettes are used for aspirating and dispensing aqueous liquids in the volume range of
0.5
–5000
μl using
GRIPTIPS® pipette tips only, see
Safety information
Regardless of the listed safety notes, all locally applicable regulations must be observed.
1) This product may only be operated in a secure, protected network with trustworthy clients.
2) The pipette may only be used by properly trained personnel in a manner specified by INTEGRA
Biosciences.
3) Use only an INTEGRA battery and charging device.
4)
Do not expose the battery to heat over 60 °C or mechanical
stress.
5) Replace battery if the charging intervals are unusually short or if the charging takes much
longer than 4 hours.
6) Do not use or charge the instrument near flammable material or in an atmosphere with danger
of explosion.
7) Servicing work and repairs may only be performed by INTEGRA Biosciences or an authorized
after-sales service member.
Battery charge
Charge the battery completely before first use.
The supply voltage must meet the requirements of the mains
adapter: 100–
240 VAC,
50/60 Hz.
Low battery
Recharging battery
Fully charged battery
~ 2.5 h
Blinking red
Blinking bars
Three solid bars
Getting Started
1.
Battery charge indicator
2.
Back button
, to navigate backward
3.
Touch wheel
, spin to scroll the cursor
4.
OK button
, to make a selection
5.
Arrow buttons
, for selections and
tip spacing (VOYAGER only)
6.
Purge button
, to empty tips
7.
RUN button
, to start operations
8.
Tip Ejector
Turn on:
Press and release
RUN
.
Turn off:
Press and hold the Back button for 3 seconds.