Electromagnetic Compatibility (EMC) User Information
(Continued)
Immunity Test
IEC 60601-1-2 Test Level
Compliance Level
Conducted RF IEC
61000-4-6
3V with 6V ISM, Amateur 80% AM
at 1 kHz or risk frequency
150 kHz – 80 MHz
V1 = 3 Vrms
Radiated RF IEC
610000-4-3
3 V/m,
80% AM AT 1 kHz or risk frequency
80 MHz – 2700 MHz
E1 = 3 Vrms
NOTE 1.
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects and people.
NOTE 3.
Table 9 testing included.
Table 3:
Guidance & Manufacturer’s Declaration – Electromagnetic Immunity
The Integra LED Surgical Headlight System is intended for use in the electromagnetic environment
specified below. The customer or user should ensure that it is used in such an environment.
28
Summary of Contents for Integra Duo
Page 1: ...Instructions for Use Integra Duo LED Surgical Headlight System ...
Page 2: ...2 ...
Page 18: ...18 ...
Page 21: ...2 Bay Charger 4 Bay Charger 21 ...
Page 33: ...This Page Left Intentionally Blank 33 ...
Page 34: ...This Page Left Intentionally Blank 34 ...