0688771-1
0688771-1
Refresher on Instructions for Use
Step 6.
With the skull clamp attached to the patient and in position, the Adaptor is brought up for
attachment to the skull clamp
Attach in this order
1. Attach swivel adaptor to skull clamp
2. Lock both locking handles
(Use an “Open-Palm” to close handles.
Make certain that the locking levers are
engaged)
3. Tighten Swivel Adaptor knob (Blue)
4. Tighten Link attachment (Yellow)
5. Double-check all connections for
being tight and secure (double-check
does not mean double-tighten)
Link torque knob
Locking Handle
Swivel/Clamp
Torque Knob
Link/Swivel
Torque Knob