29
English
Troubleshooting
Problem
Possible cause
Solution
Yellow lamp lights up
permanently.
•
Batteries are dis-
charged.
•
Charge the batter-
ies.
Liquid is not aspirated.
•
Missing or defec-
tive batteries.
•
Check that batter-
ies have been
inserted.
•
Charge the batter-
ies.
•
Replace the batter-
ies.
Liquid is not aspirated.
•
Pipettor nose piece
not tight.
•
Tighten the nose
piece.
Liquid is not aspirated.
•
The filter in the
pipettor nose piece
is wet and/or
clogged.
•
Replace the filter.