50
Unable to install the nozzle into the
InsuJet™
If you can’t push the nozzle into the
nozzle lock, it may still be in the locked
position. Please ensure that the nozzle
lock arrow points to the padlock
unlocked position, before inserting the
nozzle.
Besides the icons on the nozzle lock,
you can also look from above into the
device. You will see two triangular parts.
In unlocked position, the two parts are
aligned, allowing the nozzle with its
triangular shapes into the nozzle lock.
After the nozzle is completely pushed into
the mechanism, the second triangular part
will lock the nozzle.
UNLOCKED POSITION
LOCKED POSITION