panel of the I-1600 rack. Take the second cable, with the red marking, and connect one
end to the back of the PCI-1600, also marked with a red label, and the other end to the
connector labeled “IN Tx” on the I-1600. The connectors on either end of the cable
are keyed and will only mate when properly inserted. Do not use any cable other than
the Instrutech supplied cable. With the InstruLINK it is not necessary for the computer
or I-1600 to be power off.
Note: If connecting a second I-1600 rack unit to the PCI-
1600 then the colors connect the Green labeled cable to “IN Tx” and Yellow to “OUT
Rx”.
7. Connect the power cord to the
I-1600
. The internal power supply used in the
I-1600
is
an auto switching multi-voltage supply that will operate from 100
volts to 260 volts.
Make sure that the I-1600 power cord is
plugged into a properly grounded AC receptacle. Improper
grounding of the I-1600 could result in an electrical shock hazard.
It is advisable to plug all equipment into a common outlet strip,
this will minimize power line induced noise in the system.
Equipment plugged into different outlets can cause excessive sixty
hertz noise to be induced into the acquired data.
8. Before powering up please recheck all connections. If all connections are proper then
the power led will illuminate once the power switch is depressed.
9. You are now ready to install the acquisition software. Follow the installation
instructions provided with the software.
!