insize ISHW Operation Manual Download Page 3

5

2

  Technical specification

 Product introduction

3. anvil 

1. reading dial 

2. indenter 

Code

Application

ISHW-L20

ISHW-L20A

ISHW-L20B

Thickness requirement 

Internal diameter 

of flat workpiece

requirement of tube

workpiece

Wall thickness 

requirement of

tube workpiece

Hardness range

Min.reading

Accuracy

Dimension

for aluminum

for aluminum alloy  for aluminum 

alloy general use

thick workpiece

alloy small tube

0.6~13mm

0.6~6mm

0.6~8mm

>Ø10mm

>Ø10mm

>Ø6mm

0.6~6mm

6~13mm

0.6~8mm

(internal diameter

Ø10mm~23.3mm

0.6~6mm

(internal diameter

>Ø23.3mm)

0~20HW

1HW

±0.5HW(at 5~17HW)

205x30x85mm

4

4. pivot screw

3

20 by regulating the adjustment screw, the indenter is worn and
needs to be changed. 

2. Calibration of the load spring

Screw off the pivot screw, and take out the handle from the frame,
then remove the dial head from the indenter cylinder, leaving the
cylinder in the frame. Then the adjustment nut can be seen in the
indenter cylinder. Rotate the adjustment nut with the calibration
wrench. If the reading of the test block is smaller than the number 
marked on the test block, rotate the adjustment nut counterclockwise,
otherwise, rotate the adjustment nut clockwise. When the nut is
rotated 30o to 45o, the reading will change by 1unit. Reassemble
the tester and verify the tester both on the test block and on the
anvil after regulating it.
Note: Do not calibrate the readings of the test block by regulating
the adjustment screw.

 Changing the indenter

In case the reading can not be set to the full scale of 20 by regulating
the adjustment screw, which means the indenter is worn, the indenter
should be changed. Follow the steps below to change the indenter.
Screw off the pivot screw, and take out the handle from the frame, then
emove the dial head from the indenter cylinder, leaving the cylinder in
the frame. Then the adjustment screw can be seen in the indenter cylinder.
Rotate the adjustment screw off with the calibration wrench, then take
the indenter out and replace with a new indenter.
After the indenter is replaced, the load spring needs to be adjusted with
the adjusting nut. Only one rotation of the calibration nut is enough after
the resistance of the load spring is felt. The indenter tip will be damaged
by excessive pressure of the load spring after indenter replacement. 

Maintenance

Prevent dust and dirt.
Prevent rust.
Prevent falling off.
Do not disassemble the dial head.

)

Reviews: