insize ISHR-D121 Operation Manual Download Page 4

5

Maintenance  

Tester should NOT be stocked in carry case for a long time.

The  bottom  of  adapter  and  iron  seat  should  keep  dry  and  clean, 
resistance  of  rust.  Paint  a  little  grease  on  the  surface  of  adapter  and 
iron seat if not operating for a long time.

Do not forerake the tester, in case of damage to the screen.
Pull out the battery from box in case of contaminating the tester by its 
weeping if it is predicted not to use for some time.
It could be stocked outdoors in long time.
Do not use water or detergent to clean it.
Adapter  is  restricted  to  disassemble,  except  special  adapters  needed. 
Other  parts  are  forbidden  to  disassemble  without  instruction  and 
authority.

Notes:
a.  Test  force  should  be  held  until  time  values  disappeared,  otherwise 
the screen will display "Hold Not Enough".
b. It is better to load the total test force once, without any pause.
c. In releasing test force, "Go On" will display on screen when test force 
arrives at 0.0kg. The indenter should be upward until hardness values 
display.
d. Please change another measuring point when overload. 

Calibration

Enter calibration mode and test 1st point and get the value, then there 
will be a symbol “?” flash 6 times.

1

Now  test  2nd  point  after  the   
third point.

symbol  “?”  flash  6  times,  then  make  the 

2

The  screen  will  display  the  average  value  of  this  three  measurement, 
you  can  press  the  “STAT”  and  “CONV”  buttons  to  adjust  the  value,  at 
last press “DEL” to finish the calibration.

3