Page 44 of 48
12.
SERVICE AND MAINTENANCE
The system requires maintenance on a routine basis in accordance with SALI-D
Service Manual. The device performs periodic self-testing according to the
requirements of the Service Center. If the Service Center obtains a malfunction
indication, the Service Center will automatically alert the Client that Service is
required.
Service should only be provided by an authorized Inovytec Medical Solutions Ltd.
representative.
12.1
CLEANING AND ROUTINE MAINTENANCE
The following maintenance activities are required to be performed after each
use of the SALI-D:
Part
Procedure
Comments
SALI-D
•
Spray entire surface with 70%
Alcohol
•
Wait for 1 min.
•
Wipe device with pharma
wipes 70% alcohol
Including – head rest,
SALI-D door, and SpO
2
Sensor.
Part
Replacement
Single-use items must be
replaced:
•
Face Mask
•
ECG Electrodes
•
AED Pads
•
Oxygen Cylinder
Oxygen Cylinder needs
to be replaced after
every use or according
to local regulation.
Warning:
Do not use defibrillation pads, ECG electrodes, SpO2
Finger Probe, Face Mask, O2 Cylinder, batteries, and any other
accessory not approved by Inovytec. Use of unauthorized
accessories may cause the device to operate improperly and
provide false measurements.