INOGEN ONE G3 OXYGEN CONCENTRATOR
TECHNICAL MANUAL
Page 15 of 17
Output 19 VDC, 85 W
DC Input: 13.5 to 19 VDC, 10A
Rechargeable Battery: Voltage12.0 to 16.8 VDC,
Battery Duration:
Approximately 2 to 4 hours for 8 cell battery
Approximately 4 to 8 hours for 16 cell battery
Battery Charging
Time:
Approximately 2.25 to 4.25 hours for 8 cell
battery
Approximately 4.25 to 8.00 hours for 16 cell
battery
Environmental
Ranges Intended
for Use:
Temperature: 40 - 104
o
F (4 - 40
o
C)
Humidity: 0% to 95%, non-condensing
Classifications
Mode of Operation:
Continuous Duty
Type of Protection Against
Electrical Shock:
Class II
Degree of Protection Against
Electrical Shock:
Type BF
Not intended for cardiac
application
Degree of Protection Against
Ingress of Water:
Concentrator: IP 20
w/carry bag: IP 22
Degree of Safety for Application
in Presence of Anesthetic Gases:
Not suitable for such
applications
5.
INOGEN ONE G3 ERROR CODE TABLE
Error Code
Number
Explanation
Possible Causes &
Troubleshooting
Instructions
001
Low voltage on power
supply
Replace the power supply or
remove the battery to see
if the concentrator was
running on battery power
004
Pressure Error
The concentrator was shut
down due to a pressure
error. Check that the sieve
beds are installed
properly.
016
System electric current
out of specifications
There may be a fault with
the motor. If the error is
persistent, return the
concentrator for service