background image

 

 

Pulse Oximeter 

INV-430J/PE

 

Precautions 

  Do not attempt to repair the Pulse Oximeter by yourself. Only 

certified engineers should maintain and repair it. 

  Change the contact position between the Oximeter probe and 

the finger periodically if you are monitoring your SpO

levels 

and pulse rate for more than 2 hours.   

  This product is not designed to be used by newborn babies. 

  Seek for medical care if the measured value goes beyond the 

normal range and you are sure that the instrument is not 

malfunctioning. 

  The pulse oximeter uses infrared light (invisible to your eyes) to 

measure your SpO

2

 levels. Hence, please do not stare at the 

light-emitting components of the Oximeter, as that could cause 

harm and/or potentially blind your eyes. 

  For details about clinical limitations and contraindications, 

please carefully consult relevant medical literatures. 

The following factors may affect the accuracy of measurement: 

  The Oximeter is used in an environment involving 

high-frequency devices, such as high-frequency electric knives 

and CT apparatuses.

 

  The probe of the Oximeter is placed on the same arm that a 

blood pressure cuff arterial duct or intravenous injection. 

  The user suffers from hypotension, severe vascular atrophy, 

severe anemia, or low oxygen. 

  The user is in sudden cardiac arrest or shock state. 

  The user is wearing nail polish or artificial nails. 

Warning: 

Do not use the Oximeter in an environment with any 

flammable gases, flammable anesthetic, or other flammable substances. 

Warning: 

Keep unit and lanyard away from children as the included 

lanyard may present an entanglement or choking hazard to small 

children. Adult supervision required; never leave children unattended 

with unit or lanyard.

 

Warning:

 Do not throw the batteries into fire, as that could cause an 

explosion. 

Warning:

 Do not attempt to charge the included batteries, as that could 

cause leakage, fire disaster, or even explosion. Dispose the used 

batteries in accordance to the local laws and regulations. 

Warning:

 Do not use the Oximeter in an MRI or CT environment. 

Warning:

 

Do not operate the Oximeter if it is wet. Avoid moving the 

oximeter from a cold to a hot and humid environment. 

Warning:

 Install the batteries properly before powering on the Oximeter 

for normal use. Please remove the batteries if you are not planning to 

use the Oximeter for a long time. 

Warning:

 Close the battery cover when the instrument is in use. 

   

Symbol Conventions 

Symbol

 

Description

 

 

BF-type application part

 

 

Caution: Please see this manual. 

 

%SpO

2

 

Symbol of oxygen saturation

 

bpmPR 

Symbol of pulse rate

 

 

Temperature limitation

 

 

When end users abandon this product, they 
must send the product to the collection place 
for recycling.

 

 

Manufacturer information 

Overview 

SpO

2

 stands for peripheral capillary oxygen saturation. Oxygen 

saturation is defined as the ratio of oxyhemoglobin (HbO2) to the total 

concentration of hemoglobin (i.e. Oxyhemo reduced hemoglobin) 

present in the blood. It is an important physiological parameter involved 

in respiration and circulation. The Pulse Oximeter feature herein is small, 

portable, non-invasive and easy to use. The user only needs to insert a 

finger into the chamber to measure his/her SpO

2

 and pulse rate. 

Working Principles, Expected Usage, and Applicable Scope 

Oxygenated blood absorbs light at 660nm (red light), whereas 

deoxygenated blood absorbs light preferentially at 905nm (infra-red). A 

pulse oximeter works by passing a beam of red 

and infrared light through a pulsating capillary 

bed and then measure the amount of red and 

infrared light emerging from the tissues via a 

sensor. To improve accuracy, the NV-430J/PE 

uses a proprietary algorithm to collect data from pulsatile arterial blood 

and ignores local noise from the tissues. The relative absorption of light 

by oxyhemoglobin (HbO) and deoxyhemoglobin is then processed 

according to the Beer-Lambert's law and a qualitative measurement of 

the users’ oxyhemoglobin status i.e. Oxygen saturation level (SpO

2

derived.   

This device is intended for non-medical use in healthy people to monitor 

their pulse and blood oxygen levels for sports and/or aviation only. Do not 

use for continuous monitoring. 

Limitations 

Spot check pulse oximeter takes 1-2 reading every second and takes an 

average of 3-6 readings before displaying the result. Hence, you will 

have to wait at least 3-6 seconds before a result will be displayed on the 

monitor. If the pulse oximeter fails to detect blood flow for the first few 

seconds, then the results will be delayed accordingly.   

The pulse oximeter does not work for people with naturally small fingers 

or low peripheral blood flow. This is because the pulse oximeter would 

not have enough blood to take a measurement. Please note that prolong 

use of a pulse oximeter, hypotension, vasoconstriction, hypothermia and 

certain medications can lead to low or restricted blood flow. Nail polish 

can also impede the pulse oximeter from taking a reading.   

Finally, it is generally accepted that the saturation percentage is 

unreliable on the steep part (around 60 mm Hg) of the oxyhemoglobin 

dissociation curve.   

Innovative features of the INV-430J/PE   

The INV-430J/PE attempts to ameliorate some of the above problems by   

1) Increasing the sensitivity of the sensor so that it can measure the 

user’s SpO

2

 and Pulse Rate (PR) at blood perfusion Index (PI) as low as 

0.2%. 

2) Comes with anti-motion technology so that the pulse oximeters can 

still function even when the finger moves. 

3) Using a self-adjusting smart spring system to fit the finger snugly but 

not so tight as to impede blood flow that might lead to a low PI. 

4) Using soft, hypoallergenic medical grade silicon in the finger chamber 

so that the pulse oximeter will not cause discomfort.   

4) Using solid ABS plastic to block ambient light from reaching the sensor 

so that the signal to background ratio and the accuracy of the 

measurement is dramatically increased. 

5) Having a plethysmograph that allows the user to see and gauge the 

amount of blood detected by the pulse oximeter. This will allow the user 

to know if the finger is properly inserted into the finger chamber and if the 

pulse oximeter is taking the measurements properly (See below for 

details) 

 

Operation Guide 

Switch on the power supply of the pulse Oximeter 

Stick one finger completely into the 

finger chamber of the Oximeter. The 

fingernail should be facing upward. 

Release the clip and press the power 

key to power on the Pulse Oximeter. 

The INV-430J/PE pulse oximeter 

sensor 

Light beams   

sensor 

Reviews: