background image

__________________________________________________________________________ 

©2018 Innovasis Inc. 

Page 3 of 4 

L00022 Rev G

 

 

 

PREOPERATIVE:

  The implant components should be handled and stored carefully and protected from damage.  

The instruments and implants must be cleaned and sterilized before use.  The surgeon must confirm that all 
necessary implants and instruments are on hand for the planned surgical construct.  They should be examined for 
damage prior to use. 

INTRAOPERATIVE:

    Extreme  caution  must  be  taken  around  the  spinal  cord  and  nerve  roots,  especially  when 

inserting  screws  and  cross  connectors.    Breakage,  slippage  or  mishandling  of  the  instruments  or  implant 
components,  such  as  sharp  edges,  may  cause  injury  to  the  patient  or  operative  personnel.    The  Implants  must  be 
handled and contoured carefully so as to avoid notching or scratching the surface.  Prior to closing the soft tissues, all 
caps and screws should be tightened firmly according to the surgical technique guide.  Recheck the tightness of all 
caps  and  screws  after  finishing  ensuring  that  none  have  loosened  during  the  tightening  or  manipulation  of  other 
components. 

POSTOPERATIVE:

  The patient must be adequately instructed as to the risks and limitations of the implant as well 

as  postoperative  care  and  rehabilitation.    The  patient  should  be  instructed  in  the  limitations  of  physical  activities, 
which would place excessive stresses on the implant or cause delay in the healing process.  The patient should be 
instructed in the proper use of weight-bearing or assist devices as well as the proper methods of ambulation, climbing 
stairs, getting into/out of bed or other daily activities while minimizing rotational and bending stresses. 

The  surgeon  must  consider  the  removal  of  the  implant  after  healing  as  the  implant  can  loosen,  fracture  or  corrode 
even after fusion has occurred.  The risks and benefits of a second surgery must be carefully evaluated. 

MATERIAL  SPECIFICATION:

    Implants  are  manufactured  out  of  Titanium  6AL-4V  ELI  which  complies  with  ASTM   

F-136  or  CoCr  per  ASTM  F-1537.    INNOVASIS  expressly  warrants  that  these  devices  are  fabricated  from  material 
specifications.  No other warranties express or implied, are made.  Implied warranties of merchantability and fitness 
for a particular purpose or use are specifically excluded. 

Do  not  re-use  or  re-implant  under  any  circumstances.  Components  of  this  system  should  not  be  used  with 
components from any other manufacturer.

  

  

CLEANING AND DECONTAMINATION:

  All non-sterile implants, instruments and trays must be thoroughly cleaned 

using  neutral  cleaners  before  use  and  must  be  visibly  clean  before  sterilization  and  subsequent  introduction  into  a 
sterile surgical field or (if applicable) return of the product to INNOVASIS.   Cleaning and disinfecting of instruments 
can be performed with aldehyde-free solvents at higher temperatures.  Cleaning and decontamination must include 
the use of neutral cleaners followed by deionized water rinse. 

Note:    certain  cleaning  solutions  such  as  those  containing  formalin,  gluteraldehyde,  bleach  and/or  other 
alkaline cleaners may damage some devices, particularly instruments; these solutions should NOT be used.  
Also, many instruments require disassembly before cleaning. 

Detailed cleaning recommendations for reusable instruments are in the Innovasis 

Surgical Instrument Reprocessing 

Instructions

 LG01.  A copy of this guide is available at 

www.innovasis.com

 or by request via telephone, fax or e-mail.  

Contact information appears at the heading of this document. 

All  products  should  be  treated  with  care.    Improper  use  or  handling  may  lead  to  damage  and/or  possible  improper 
functioning of the device. 

STERILIZATION:

  

The INNOVASIS

®

 EXCELLA

®

 SPINAL SYSTEM implants are supplied

   

NON-STERILE

.  The 

metal implants and instruments used in surgery MUST be sterilized by the hospital prior to use.  FDA cleared wraps 
are recommended for use with the sterilization tray.  Remove all packaging materials prior to sterilization.  Only sterile 
products should be placed in the operative field.  For a 10

-6

  Sterility Assurance Level, it is recommended that the 

Excella

®

 Spinal System implants be steam sterilized by the hospital using one of the two sets of process parameters 

below: 

 

METHOD 

CYCLE 

TEMPERATURE 

EXPOSURE  TIME 

DRY TIME 

Steam 

Pre-vacuum (Wrapped) 

270

0

 F  (132

0

 C ) 

4 Minutes 

60 Minutes* 

Steam 

Gravity Displacement 

(Wrapped) 

270

0

 F  (132

0

 C ) 

30 Minutes 

60 Minutes* 

 

*This dry time sterilization cycle (60 minutes) is not considered by the Food and Drug Administration to be a standard 
sterilization cycle.   
It  is  the  end  user’s  responsibility  to  use  only  sterilizers  and  accessories  (such  as  sterilization  wraps,  sterilization 
pouches,  chemical  indicators,  biological  indicators,  and  sterilization  cassettes)  that  have  been  cleared  by  the  Food 
and Drug Administration for the selected sterilization cycle specifications (time and temperature).

 

NOTE:    Because  of  the  many  variables  involved  in  sterilization,  each  medical  facility  should  calibrate  and 
verify the sterilization process (e.g. temperatures, times) used for their equipment. 

CAUTION:  Federal Law (U.S.A.) restricts these devices to sale by or on order of a physician.

  

 

Summary of Contents for Excella III-D

Page 1: ...nct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic lumbar and sacral spine Degenerative disc disease DDD defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies Degenerative spondylolisthesis with objective evidence of neurologic impairment Scoliosis Fracture or dislocation Kyphosis Spin...

Page 2: ... site pain ii Nerve or vascular damage due to surgical trauma x Loss of fixation iii Mal alignment of anatomical structures xi Infection iv Hemorrhage of the blood vessels and or hematomas xii Bursitis v Gastrointestinal urological and or reproductive system xiii Non union or delayed union compromise including sterility impotency and or loss of xiv Bone loss due to resorption or stress consortium ...

Page 3: ...INATION All non sterile implants instruments and trays must be thoroughly cleaned using neutral cleaners before use and must be visibly clean before sterilization and subsequent introduction into a sterile surgical field or if applicable return of the product to INNOVASIS Cleaning and disinfecting of instruments can be performed with aldehyde free solvents at higher temperatures Cleaning and decon...

Page 4: ...ge 4 of 4 L00022 Rev G Symbol Glossary hCatalog Number gBatch Code dNon Sterile Do not re use Consult Instructions for Use MManufacturer Rx Only Prescription only CoCr Cobalt Chrome Ti 6Al 4V ELI Titanium Alloy Extra Low Interstitial 316 Stainless Steel 316LVM Stainless Steel Stainless Steel Alloy ...

Reviews: