Concentrating Pipette

 Quick Start Guide 

                                                (US Patents 8,110,112, 8,584,535, D688,805

 

 and other patents pending) 

 

 

Setting up the instrument: 

1.

 

Plug the 24VDC power supply into the 

24VDC Jack

 on the side of the instrument

.

 

2.

 

Attach the permeate tube to the 

Permeate Port

 using the quick connect fitting and insert the other end of the permeate 

tube in a waste container. 

3.

 

Flip the 

Power Switch

 to turn the instrument on, it will take approximately 5 seconds for the unit to power up. 

4.

 

Press enter to advance from the welcome screen. 

5.

 

If desired, the user can ensure that the instrument was not damaged in shipping by performing the Concentrating Pipette 
Check List contained in Section 6.7 of the Concentrating Pipette User’s Guide. 

6.

 

To decontaminate the unit after receipt perform the following: 
a.

 

Insert the 

Maintenance Tip

 received with the unit into the 

Tip Interface Ports

 located on the underside of the 

Fluidics 

Head

. The larger port on the 

Maintenance Tip

 should be at the back (the 

Maintenance Tip

 can only be inserted one 

way). You should feel the 

Maintenance Tip

 “click” into place, and you should be able to confirm that the 

Maintenance 

Tip

 is fully seated by looking through the clear plastic of the 

Fluidics Head

b.

 

Select the following on the 

Control Panel

 “Options”>”Procedures”>”Decon cycle”.  

c.

 

Install the Decontamination Fluid Can from the Maintenance Fluid Kit by first placing the tip of the inverted can into the 

Can Interface Port

 on the right side of the instrument. Then push the can down and latch it under the 

Can Latch

 

d.

 

Select “Start” by pushing enter.  The unit will count down from 10 minutes. 

e.

 

The unit will automatically perform a flush of the system. 

f.

 

Install the Rinse Fluid Can from the Maintenance Fluid Kit into the 

Can Interface Port

 as above

.

 

g.

 

Select “Start” by pushing enter.   

7.

 

Install an elution 

Fluid Can

 into the 

Can Interface Port

 as above.  

8.

 

It is necessary to purge air from the extraction fluid line and prime the fluidics when a new fluid can is installed 
or when the instrument has been idle with an elution fluid can installed for a period of 4 hours or overnight. The 
quality of the foam may be reduced due to off‐gassing of CO

in the fluid lines. To purge the fluid lines: 

a.

 

Hold a waste container under the 

Tip Interface Ports

 on the bottom of the 

Fluidics Head

.  

b.

 

Select the following on the 

Control Panel

 “Options”>”Procedures”>”Ext. fluid prime”. A small amount of foam will come 

out of the front 

Tip Interface Port

9.

 

Select “Return” to go back to the options menu, select “Ext. size” and choose the desired final extraction volume. Further 
control of the extraction volume is found under the “Adv. Options” menu. 

Concentrating a sample: 

1.

 

Insert a 

Single-use Concentrating Pipette Tip

 into the 

Tip Interface Ports

 located on the underside of the 

Fluidics Head

.  

2.

 

Place the sample on the 

Sample Platform

 under the 

Single-use Concentrating Pipette Tip

 and lower the 

Fluidics Head

 so the 

Single-use Concentrating Pipette Tip

 reaches all the way to the bottom of the sample container. Conical sample containers 

are recommended. 

3.

 

From the main menu, select “Start run”. The pump will start, a red light in the 

Fluidics Head

 will flash and the 

Control Panel

 

will display “Running” with a timer. It will take several seconds for the liquid to reach the pump, once it does it may sound 
louder. 

4.

 

When the entire sample has been pulled through the 

Single-use Concentrating Pipette Tip

, the unit will stop automatically. 

This will occur 3 to 30 seconds after the last of the fluid is pulled into the tip. 

5.

 

When the run is finished, raise the 

Single-use Concentrating Pipette Tip

 out of the sample container. Hold a small sample 

container under the 

Single-use Concentrating Pipette Tip

 and select “Extract”. The extraction cycle will take about 10 

seconds and a jet of white foam will be eluted from the 

Single-use Concentrating Pipette Tip

 into your sample container. 

6.

 

If you run out of elution fluid during an extraction, remove the concentrating tip, replace the elution can, perform an Ext. 
fluid prime, re-insert the concentrating tip, perform another extract. 

7.

 

Remove the used 

Single-use Concentrating Pipette Tip

 by pushing your thumb up against the bottom of the 

Fluidics Head

 

while you pull the 

Single-use Concentrating Pipette Tip

 down with your hand, the 

Single-use Concentrating Pipette Tip

 is 

only held in place by a detent. 

 

 