Innova ALPHAiX Series Operating Instructions Manual Download Page 9

APPENDIX

Explanatory Notes on Using the Measuring Table

As can be seen, the relationship between the impulses of the counter tubes is proportionate to 
the becquerel values - in other words, the higher impulse rates mean correspondingly higher 
becquerel values. Conclusions for other measurements are thus possible.

If, for example, a specific object contaminated by caesium-137

***

 must be examined, a 10 

minute measurement at a distance of 30 mm from the sample should be carried out. The result 
reduced to 1 minute must then be used for the table.

EXAMPLE:

A 10 minute measurement on a sample with caesium-137 using a Type G
counter tube shows after the measuring time a measured value of 500 impulses.

After reduction to 1 minute (500 : 10 = 50 Ipm) and after deduction of the 
background counts (30 Ipm) a net impulse rate of 20 Ipm is left.

The column in the measuring table for a counter tube G shows under 100 Bq 
Cs-137: 143 Ipm. Consequently, 20 Ipm corresponds to 

100 Bq / 143 x 20 = 14 Bq

(14 Bq x 50 = 700 Bq/kg)

Experience shows that the pre-requisites often do not concur with those of the measuring 
table. A shorter distance, usually 5 mm, is often chosen for surface measurements using the 
type A or G end window counter tubes. The number of impulses at a distance of 5 mm is 5 
times as high as that shown in the table, i.e. the corresponding value in the measuring table 
must be multiplied by a factor of 5 before conversion. 

Thus 715 Ipm would correspond to (143 x 5) 100 Bq. 
Converted to the above 30 Ipm, that would be 

(100 / 715 x 20) only 2.8 Bq Cs-137

The immersion counter tubes are not normally used for surface measurements. These are much 
more e

ffi

cient as immersion probes. To obtain comparable results, the value in the measuring 

table must be multiplied by the high factor of 10 in this case. This means that 100 Bq Cs-137 

*** It must be assumed that the existing contamination in Europe due to the Chernobyl disaster 
must be attributed almost exclusively to the nuclide caesium-137.

OPERATING INSTRUCTIONS 

for Counter Tube Type G

JS Monday 10 January 2011                                                                                                                                                  Page  9 of  9

If the sample weight is, for example, 20 gram this value must be extrapolated to 1 kg 

would correspond to 1430 Ipm (143 x 10) for a counter tube type G.

Summary of Contents for ALPHAiX Series

Page 1: ...fferentiation of radiation types Page 6 7 Some more theory Page 7 APPENDIX Measuring table Page 8 Using the measuring table Page 9 ALPHAiX with Counter Tube Type G End window counter tube for measurem...

Page 2: ...active radiation source present in its vicinity The recorded number of impulses is converted to a value per minute This value Ipm is then the so called background count In measurement only measured va...

Page 3: ...to contact For accurate measurements a stand is required to ensure a constant distance for 10 minutes As already mentioned the tolerance value for a 10 minute measurement is 35 Ipm i e if a maximum of...

Page 4: ...ust be extended if a 10 minute measurement does not show satisfactory results As you can see we have rounded up in the calculation of the tolerance value Needless to say you can work with the more acc...

Page 5: ...with opened covers both at the tablet and the counter tube with the end window of the tube held directly on the tablet surface In the counter tube A the metal bezel can be put on the tablet as the en...

Page 6: ...the two measurements If ALPHA radiation is present the first measurement without transparent film must be correspondingly higher The distance for these measurements should be 5 mm Separating the BETA...

Page 7: ...tivity Bq of the radiation source required to allow its measurement The radiation energy keV and its activity Bq are two different factors which together with the type of radiation ALPHA BETA and GAMM...

Page 8: ...n for this measurement Smaller distances give higher impulse frequencies and larger distances correspondingly lower counting efficiencies Impulses per minute Ipm NUCLIDE END WINDOW COUNTER TUBES IMMER...

Page 9: ...ence shows that the pre requisites often do not concur with those of the measuring table A shorter distance usually 5 mm is often chosen for surface measurements using the type A or G end window count...

Reviews: