4.0 General Operation

 

 

Note:

   See Demo Video on www.YouTube.com 

 

(Search SpotOn Sprayer Calibrator) 

 

4.1 Taking a Measurement 
 

 

Assure batteries have been installed in the 
meter.  Press the 

START

 button to turn the 

meter on.  The LCD will flash “000” 
momentarily indicating it is working properly.  
The LCD will then show a “_” marker to 
indicate the type of measurement units the 
meter will use when displaying readings.  The 
meter will then show “- - -“ on the LCD 
indicating the meter is ready to take a new 
measurement.  Once the meter shows “- - -“, it 
can be placed under the spray nozzle to be 
tested (you have 60 seconds to place the 
meter under the nozzle).  Placement of the 
meter under the nozzle should be brisk and 
certain so that all the flow from the nozzle is 
transferred quickly to the meter (See Fig 1).  
As the meter fills with water, the display will 
start flashing “- - -“; this indicates that the 
reading is in progress.  As soon as water 
reaches the meter’s upper sensor, the flashing 
stops and the flow rate is displayed on the 

LCD (See Fig 2).  This flow rate 
will continue to be displayed on 
the meter for 90 seconds or until 
the 

START

 button is pressed to 

start a new reading.  Once the 
reading has been displayed, the 
water can be poured out of the 
open top of the meter (See Fig 
3).  Pouring out the contents of 
the meter will not affect the 

reading being displayed on the LCD.  If a reading must be re-started 
for some reason, simply empty the meter and re-press the 

START

 

button to ready it for a new measurement.   

 

 

Fig. 1 

 

 

Fig. 2 

 

Fig. 3 

 

5

PLEASE NOTE:  The most accurate measurements are made with 
the meter held at a slight angle to vertical as seen in Fig. 1.  This 
keeps the water flowing down into the meter along the back side of 
the meter and keeps it from accidentally falling on and triggering the 
meter’s sensors. 
 

4.2 Turning the Meter Off 
 

The meter will turn on any time the 

START

 button is pressed.  The 

meter can be turned off in two ways:   

 

1.   The meter can be shut off by pressing and holding the 

START

 

button for several seconds (about 3 seconds).  As soon as the 
LCD goes blank, release the START button.   

 

2.  The meter will automatically shut itself off after 90 seconds of 

inactivity.  This means that the user does not have to shut off 
the meter after use since it will eventually shut itself off. 

 
4.3 Changing Measurement Units 
 

The meter can be configured to display readings in either gallons per 
minute (gpm), ounces per minute (oz/min), or liters per minute 
(L/min).  To change the display units, the user must first turn the 
meter off.  Then press and hold the 

START

 button.  Keep holding the 

START

 button until the display shows “_ _ _” (about 3 seconds).   

When the display shows “_ _ _” release the 

START

 button and the 

display will start flashing “_” above the unit type that is currently 
selected.  Press the 

START

 button to change this selection.  Once 

the correct unit type is selected, wait without pressing the 

START

 

button for 5 seconds.  The meter will then enter the new setting and 
shut off.  From that point forward, the meter will always display 
readings in the new units. 

 

 
 
 
 
 
 
 

Summary of Contents for SpotOn SC-4

Page 1: ... 2 Features 2 2 0 Specifications 3 3 0 CE Declaration 3 4 0 General Meter Operation 4 4 1 Taking a Measurement 4 4 2 Turning the Meter Off 5 4 3 Changing Measurement Units 5 4 4 Installing the Batteries 6 5 0 Trouble Shooting 6 6 0 Warranty Service 7 7 0 French Instructions Instructions français 8 ...

Page 2: ...z min 0 01 L min to 9 99 or 0 1 L min 10 0 up Battery two 2 N size alkaline batteries Battery Life 2 years meter has low battery indicator Size 10 0 in 25 4cm tall by 4 0 in 10 2cm diameter Weight 1 2 lbs 544 g weight Construction Break resistant plastic stainless steel Country of Origin Made in USA 3 0 Declarations CE Declaration of Conformity Innoquest Inc 910 Hobe Road Woodstock IL 60098 USA Ef...

Page 3: ... meter and re press the START button to ready it for a new measurement Fig 1 Fig 2 Fig 3 5 PLEASE NOTE The most accurate measurements are made with the meter held at a slight angle to vertical as seen in Fig 1 This keeps the water flowing down into the meter along the back side of the meter and keeps it from accidentally falling on and triggering the meter s sensors 4 2 Turning the Meter Off The m...

Page 4: ...bility to sense the water level Care should be taken to rinse the meter inside and out with clean tap water when measurements are complete This will keep any foreign material from building up on the electrodes Complete rinsing will also keep the diffuser pad from becoming clogged and therefore hindering the flow of water into or out of the meter during operation Fig 4 7 6 0 Warranty Service One Ye...

Page 5: ...rrer Fig 1 Fig 2 Fig 3 9 VEUILLER NOTER les mesures les plus précises sont réalisées avec le calibreur maintenu avec un léger angle de la verticale comme sur la Fig 1 Ceci permet au flux de s écouler plus lentement dans I appareil et empêche tout déclenchement intempestif des capteurs du calibreur Arrêter le calibreur L appareil sera mis en route chaque fois que l on presse sur la touche START Le ...

Page 6: ...oivent être remplacées comme Problèmes de fonctionnement Le calibreur mesure le débit en comptant le temps nécessaire au remplissage du tube II y a 3 électrodes dans le tube qui décèlent l élévation du niveau de I eau Le calibreur ne fonctionne qu avec des fluides qui conduisent le courant comme par exemple I eau Dans le cas d utilisation d eau distillée celle ci présente une trop faible conductiv...

Reviews: