View950M Multi-Scan
INNO Instrument Inc. Page 45 of 55
<
Connection >
<
Progress >
<
Completion >
3.
OSL – Load Calibration
- When the short calibration is completed, the screen switches to the
Load connection.
- Remove the short cal kit connected to the real port and connect
the load.
- After connecting the load cal-kit to the port, click the Load Icon as
shown on the left.
- Load calibration is performed as below.
<
Connection >
<
Progress >
<
Completion >
4.
Current Port Calibration completion
- After completing Open, Short, and Load, the corresponding icon
on the screen will change to
, and the 'Apply' button will appear
on the bottom button array.
- Click the 'Apply' button to apply the actual calibration.