4. MAINTENANCE

 

4.1  Apparatus maintenance

 

Do not leave the evaporator surface contaminated from corrosive substances. It could damage the 

evaporator coating. Clean the contamined surface with a clean soft cloth. The cloth may be moistened 

but not wet. It is also possible to use ordinary washing means for cleaning purposes. It is forbidden to 

clean  the  heating  bath  with  the  means  which  could  spoil  its  smooth  surface.  In  case  of  sediments  of 

furning,impurities, particulates of water rusts  ,use the cleaning compounds for cleaning of furning or 

mellow solution of HCL. In process of cleaning use the protective means recommended  by producer of 

applied cleaning compounds.  

Caution: Disconnect the apparatus from the network while cleaning it with a wet cloth!

 

4.2  Sealing replacement

 

       Disassemble  the  condenser,  remove  the  releasing  pipe  with  the  valve,  remove  the  ball-and-

socket  ground  joint  flask,  release  the  nut  of  the  ball  adapter  and  carefully  remove  the  actual  adapter. 

Release the gufero sealing, clean the ball adapter, slightly apply silicone Vaseline onto the new sealing 

and mount it into the ball adapter. The sealing cavity faces out of the ball adapter. If necessary, replace 

also  the  flat  sealing  between  the  ball  adapter  and  the  head  fitted  with  a  rotary  casing.  Perform  the 

assembly  according  to  the  point  3.2.  Apply  a  light  layer  of  silicone  Vaseline  to  the  gufero  sealing 

approximately after 80 operation hours (more often if necessary). 

 

5. DEFECTS AND TROUBLESHOOTING 

5.1  Leakage 

The main cause of leakage is usually an incorrect sealing assembly. At first you should check that 

all joints have been installed correctly. If it is the case, disassemble the glass and verify the condition of 

the ball adapter sealing and flat sealing, if necessary replace them. Carefully assemble according to the 

instructions, switch the rotation ON and start the vacuum pressure up.

 

5.2  Other failures 

The instrument does not react after being turned on. Check the fuse or replace for a new one 

of the same parameters, as appropriate. Check that the rotation is not hindered (turning by hand), 

turn on and off the network switch. 

The  window  of  the  rotation  key  is  flashing,  rotating  parts  are  not  moving.  Ensure  that  Rotation 

blocking knob is set to ”Unlock” and than press rotation key again.

 

The window of the rotation key is 

flashing.  Check  that  there  is  not  anything  which  could  impede  the  flask  from  its  rotation  (manually), 

switch the mains switch OFF and ON again.

  

The bath no warms. Put out the socket plug of the bath and let the bath cool down. Plug in again fill 

up the water or oil and set required temperature. 

 

 

 

 

Summary of Contents for RVO 004

Page 1: ...INGOS Ltd Laboratory Instruments division Page 1 INGOS Ltd ROTARY VACUUM EVAPORATOR RVO 004 User manual PRAGUE July 2018...

Page 2: ...e 2 Producer Laboratory Instruments division of INGOS Ltd Supplier and service INGOS Ltd Tel 420 225 983 400 K Nouzovu 2090 420 225 983 409 143 00 PRAGUE 4 Fax 420 244 403 051 CZECH REP e mail instrum...

Page 3: ...INGOS Ltd Laboratory Instruments division Page 3...

Page 4: ...dense in a vertical cooler and are collected in a flask This ensures that the evaporated solution cannot be damaged and therefore impaired RVO 004 is equiped by water bath with the electronic temperat...

Page 5: ...lve flexible and fixed clamp For complete list of basic accessories see paragraph 7 1 1 Fitting Vacuum connection 8 Stand 2 Fitting Cooling water connection 9 Rotation controling 3 Condenser cooler 10...

Page 6: ...ted in the housing and carefully insert the ball adapter with Gufer and the nut on glass pipe 7 Insert the ball adapter until the end stop and tighten it carefully with nut 2 The half ball joint of th...

Page 7: ...INGOS Ltd Laboratory Instruments division Page 7 1 Ball adapter 5 Flat sealing 2 Nut 6 Housing 3 Bead type securing ring 7 Suck pipe 4 Gufero sealing 35 22 10 Fig 2a Ball adapter sealing...

Page 8: ...al resistant 3 3 Apparatus control The instrument is put into operation by connecting the network switch The round bottomed flask starts rotation after turning on the switch The speed of rotation is c...

Page 9: ...eplace also the flat sealing between the ball adapter and the head fitted with a rotary casing Perform the assembly according to the point 3 2 Apply a light layer of silicone Vaseline to the gufero se...

Page 10: ...er increasing the concentration of the evaporated sample If the flask cannot be removed easily from the cone carry out further operation only after cooling the bath Caution If you are replacing the ro...

Page 11: ...ving flask 500 ml joint KS35 20 4SKL0006 Receiving flask 1000 ml joint KS35 20 4SKL0034 Receiving flask 2000 ml joint KS35 20 4SKL0005 Vertical condenser 4SKL0033 Diagonal condenser 4SKL0020 Dry ice t...

Page 12: ...4 Flat sealing 1TOO0010 Scraper ring 1TOO0008 Sealing PTFE Fig 2b RO1080 Fixed clip for KS32 20 RO1041 Elastic clip for NZ14 23 4SKL0026 Straight fitting GL 14 4SKL0027 Cap nut GL 14 90000013 Tubing c...

Page 13: ...INGOS Ltd Laboratory Instruments division Page 13 Fig 3 Glass accessories overview...

Page 14: ...it over to the buyer The instrument may be used only in the manner described in these instructions for use The instrument must not be used in a manner different from that indicated in these instructi...

Page 15: ...10 5 1 Leak required pressure is not obtained 10 5 2 Other failures 10 6 SAFETY AND OPERATION CONDITIONS 11 6 1 Safe operation 11 6 2 Operating conditions 11 7 ACCESSORIES AND SPARE PARTS 11 7 1 Basic...

Reviews: