Infopia EasyGluco Operation Manual Download Page 18

[ 27 ]    

EasyGluco

TM

Troubleshooting

Please refer to the table below to identify problems that persist with no messages appearing.

PROBLEM

CAUSED BY

WHAT TO DO

The meter does not
power on

There may be a problem with
the test strip

Re-test with a new strip.

Re-code the meter and re-test.

The codes of the meter and
of the test strip vial do not
match

Change the batteries and if the problem persists, contact
your local representative.

Batteries are dead or there is
a problem with the meter.

Poor amount of sample

Apply a sufficient amount of the blood sample to a new test
strip and re-test.
Re-test with a new strip.

Problem with the test strip

Test does not start
after applying blood
sample

The result is doubtful

INFOPIA

[ 28 ]

Specifications

Model   

IGM-0002A                                                  

Test Range  

0.6~33.3 mmol/L (10~600 mg/dL)

Reading Time   

9 sec

Memory Capacity   

200 test results

Operating Temperature 

50-104˚F(10-40˚C )

Operating Humidity

10 ~ 90%

Sample Type  

Capillary Whole Blood

Calibration

Plasma-Equivalent

Hematocrit

30 ~ 55%

Sample Volume   

Minimum 1.5 

Display Type    

LCD

Size (D X W X H)  

22.8 

54.5 

X  

91 (mm)

Weight  

55g (including batteries)

Power Source  

3V Li Battery (CR2032) 

2

Battery Life  

running 5000 tests 

Glucose Unit

Either mmol/L or mg/dL

Specifications

Summary of Contents for EasyGluco

Page 1: ...80 Korea www infopia21 com ObelisS A Bd GeneralWahis 53 1030 Brussels Belgium Tel 32 2 732 59 54 Fax 32 2 732 60 03 E Mail mail obelis net Lancing Device GMMC Room No 1112 AceTower 9th Bldg 345 30 Kas...

Page 2: ...Blood Glucose Monitoring System Operation Manual EasyGlucoTM...

Page 3: ...with your healthcare professional before making any changes The EasyGlucoTM System is for in vitro diagnostic use only A warranty registration card is included with your system Please complete the thr...

Page 4: ...ontact your local representative Before using the EasyGlucoTM Blood Glucose Monitoring System read all of the operating instructions in this Operation Manual to practice forsafeandaccuratetesting Easy...

Page 5: ...r system to the place of purchase or contact your nearest Infopia representative EasyGlucoTM Blood Glucose Monitoring System KIT Contents EasyGlucoTM Blood Glucose Meter Test port Product Label Batter...

Page 6: ...se Meter Display LCD display segment System Check Unit of result EasyGlucoTM Blood Glucose Test Strip The EasyGlucoTM System measures the amount of glucose in your blood Apply blood here Check here to...

Page 7: ...ormation Store the EasyGlucoTM Test Strip vials in a cool dry place between 36 86 F 2 30 Keep out of direct sunlight Do not freeze Store test strips in it s original vials only Do not mix the test str...

Page 8: ...o obtain accurate results The control solution test is similar to a blood test except that you use the EasyGlucoTM Control Solution instead of a drop of blood Code numbers are used to calibrate the Ea...

Page 9: ...on test STEP 1 Insert Test Strip Insert a Test Strip with the arrow side facing up and in the direction of the arrow insert it into the test port of the Meter firmly The meter will automatically turn...

Page 10: ...EasyGlucoTM Control Solution 11 EasyGlucoTM INFOPIA 12 The EasyGlucoTM Meter has three functional modes as settings 1 Average period 2 Unit 3 Date Time If you need to change the time or if you remove...

Page 11: ...ton down With the correct minute on display press the MEM button and the 12 hour format AM PM setting will start flashing Press the SET button to select the appropriate time format Press the MEM butto...

Page 12: ...ck on the lancing device Discard the used lancet to a proper place Clean the lancing device if blood or stain remains CAUTION Use a new sterile lancet every time you test Do not use lancet if protecti...

Page 13: ...procedure from the beginning If the codes on the LCD display window and on the test strip vial do not match do the Coding The Meter procedure for accurate result Practice using the Lancing Device and...

Page 14: ...syGlucoTM Meter CAUTION If you receive any error messages please refer to the Troubleshooting section of this Manual The EasyGlucoTM Meter saves up to 200 test results in built in memory and records t...

Page 15: ...revent infectious diseases 2 The meter should be cared for after testing and cleaned if necessary with a soft cloth or tissue If necessary rubbing alcohol can be used to wipe away dirt on the outer su...

Page 16: ...s with two 3 0V CR2032 or equivalent lithium batteries The batteries will provide you with enough power to perform about 5 000 tests When replacing the battery 3 0V CR2032 or equivalent lithium batter...

Page 17: ...ution With normal result re test with blood samples two or three times HI results indicate severe hyperglycemia high blood glucose If the HI message persists please seek medical attention immediately...

Page 18: ...ly a sufficient amount of the blood sample to a new test strip and re test Re test with a new strip Problem with the test strip Test does not start after applying blood sample The result is doubtful I...

Page 19: ...batteries supplied with the Meter set Please fill in the warranty card attached to the set Mail or fax it back to us The EasyGlucoTM Meter has a full three year warranty from the original date of your...

Page 20: ...Memo Memo...

Reviews: