Type N Coaxial Calibration Kits 

 

 

4

 

   

 

 

 

 

 

 

 

        

    

fairviewmicrowave.com

 

 

General Information 

Calibration Kit Description 

These  Fairview Microwave  N  coaxial calibration kits are designed to provide accurate calibrations of 

network analyzers in the DC to 18 GHz range. These kits include all the necessary calibration standards 

and associated hardware needed for the accurate calibration of most network analyzers. This manual 

applies to the following model numbers: FMCK1025 8-piece SOL (Short, Open, Load) calibration kit and 

FMCK1026 11-piece SOLT (Short, Open, Load, Through) calibration kit.  
Refer to the 

Calibration Kit Contents

  section  for information on included components and available kit 

options. 
NOTE: This document, along with the kit data file and datasheet, can be downloaded from 

fairviewmicrowave.com. 
 

Maintenance 

This calibration kit is relatively maintenance-free, if the components are handled with the same care that is 

appropriate to all precision equipment. As with any precision component, proper care should be taken to 

ensure clean mating surfaces, correct alignment when mating, and proper torqueing of connectors. To help 

maintain the integrity of the components in the kit, routine visual inspection and cleaning of mating surfaces 

is recommended. Failure to do so may result in degraded repeatability and accuracy, as well as damage to 

any mated devices. 
 

Calibration 

To maintain and certify the calibration kit’s ongoing performance to specification, we recommend that all 

kits be periodically returned to Fairview Microwave for calibration. The typical calibration cycle is one year, 

although actual needs may vary depending on usage.  
 

Supporting Test Port Adapters 

When configuring a test setup, ensure that damaging stresses are not applied to the connectors on the test 

set. This is particularly critical when the attached components are heavy or long. Always properly support 

the test port adapters being used. 
 

Electrostatic Discharge Precautions 

Protection against electrostatic discharge (ESD) is essential while inspecting, cleaning, or making 

connections to connectors attached to a static-sensitive circuit, such as those found inside test sets. 
When handling the connectors on the test set, be aware that you are coming into contact with exposed 

center conductors that are connected directly to the static-sensitive internal circuits of the network analyzer. 

Ensure that you and your equipment are well-grounded before inspecting, cleaning, or making connections 

to test set ports. Standard ESD precautions, such as the use of grounded wrist straps and grounded 

antistatic mats, are recommended.

 

 

Summary of Contents for Fairview Microwave FMCK1025

Page 1: ...Electronics Inc User Manual N Coaxial Calibration Kit DC to 18 GHz Models FMCK1025 FMCK1026 Fairview Microwave 301 Leora Ln Suite 100 Lewisville TX 75056 Tel 1 800 715 4396 or 972 649 6678 Fax 972 64...

Page 2: ...t be returned to Fairview Microwave and must be issued a return authorization number by Fairview Microwave prior to shipment Buyer shall prepay shipping charges to Fairview Microwave Obligation is lim...

Page 3: ...nty 2 General Information 4 Calibration Kit Description 4 Maintenance 4 Calibration 4 Supporting Test Port Adapters 4 Electrostatic Discharge Precautions 4 Connector Description 5 Connector Care 5 Con...

Page 4: ...pection and cleaning of mating surfaces is recommended Failure to do so may result in degraded repeatability and accuracy as well as damage to any mated devices Calibration To maintain and certify the...

Page 5: ...er connector contact pin depths should also be verified through regular inspections using a connector gage kit to ensure that connectors on both calibration devices and devices under test DUTs have co...

Page 6: ...stationary while torquing the nut on the other device or cable This open end wrench is supplied with the calibration kit for this purpose Using the torque wrench Hand tighten the connection being torq...

Page 7: ...ea Open male FMSC3039 1 ea Load female FMTR1068 1 ea Load male FMTR1069 1 ea 9 16 Open End Wrench FMTL1004 1 ea 13 16 Torque Wrench ST N 1316 BO14 1 ea Short female FMSC3023 1 ea Short male FMSC3024...

Page 8: ...ments made by the instrument The calibration standards in this kit have precisely known electrical behavior and during calibration the VNA software uses the raw measurement data and the known behavior...

Page 9: ...2020 Infinite Electronics Inc Fairview Microwave is a registered trademark of Infinite Electronics Inc 9...

Page 10: ...ttps www fairviewmicrowave com nsearch aspx Category Calibration Kits Portable Calibration Kits Calibration Kits Components keywords calibration ports vna sort y searchtype 1 view_type grid Fairview M...

Page 11: ...ics Inc 11 Contacts Customer Support Sales Fairview Microwave 301 Leora Ln Suite 100 Lewisville TX 75056 USA Phone 800 715 4396 972 649 6678 Fax 972 649 6689 Sales Email sales fairviewmicrowave com Cu...

Reviews: