5 - 34
XTC/3 Operating Manual
5.2.2.6 Standard REMOTE Commands
R <Command ID> (<Action Value>)
Command ID = <Byte>. See "R Command ID" column in
Action Value = <Byte> | <Float>
Float = 4 byte, ANSII standard, single precision, low to high
A few commands have a value with them. See
.
Response = None (Just header and trailer)
Table 5-5 Standard remote commands
R
(0x52)
Command ID
Function
Description
0 (0x00)
Start
Starts the next layer of the active process.
1
(0x01)
Stop
Stops the layer, puts power to zero,
closes shutters, etc.
2
(0x02)
Reset
If in Stop, resets XTC/3 to the Ready state,
at the beginning of the process.
3
(0x03)
Remote lock on
Prevents the update of parameters via
the front panel.
4 (0x04)
Remote lock off
Removes the remote lock.
5
(0x05)
Crystal fail inhibit on
Prohibits the crystal fail relay from activating.
6 (0x06)
Crystal fail inhibit off
The crystal fail relay will act normally.
7
(0x07)
Soak Hold 2 on
Will cause the unit to stay in Soak 2 state,
once entered, until the Soak Hold is removed
(R8)
8 (0x08)
Soak Hold 2 off
Will allow the unit to leave the Soak 2 state.
9 (0x09)
Manual on
Places the unit in the Manual power state.
10
(0x0a)
Manual off
Removes the unit from Manual power state,
and puts it into the Deposit state.
11
(0x0b)
Set power to <Value> vv
Active layer to vv % power.
vv = <Float> 0 - 100 (Note: Only allowed if
current state is Manual)