6 | Operation
INFICON
72 / 140
UL3000-Fab-Operating-instructions-iina73en1-08-(1910)
ð
You can change the following settings, to view the preset values see “Factory
31]“.
“Pressure SL200 capillary blocked“ (lower limit)
“Pressure SL200 capillary broken“ (upper limit)
“Pressure SL3000 capillary blocked“ (lower limit, UL3000 Fab PLUS only)
“Pressure SL3000 capillary broken“ (upper limit, UL3000 Fab PLUS only)
2
Set up.
3
Save
.
6.3.16 Calibrating
In order to be able to measure leak rates correctly, the leak detector has to be
adjusted at regular intervals by calibration. The mass spectrometer is adjusted
automatically.
You can perform the calibration using the device internal calibration leak or using an
external calibration leak. Normally, calibration should take place on a daily basis or if
there is some doubt as to the measurement ability of the device.
► When using hydrogen or helium-3 (
3
He) as tracer gas an internal calibration is not
possible. In this case, use an external calibration leak.
► The operation modes "Vacuum" and "Sniffing" must be separately calibrated when
using an external calibration.
► A calibration should only take place when the device is at operating temperature.
Calibrate at least 20 minutes after switch-on.
► For systems with a greater distance to the test object, we recommend the use of an
external calibration leak at the location of the test object.
► When using the internal calibration leak, observe the setting of the machine factor,
see also “Setting the machine factor [
65]“.
6.3.16.1 Internal calibration
ü
Operator or Supervisor rights
ü
The device is in the state "Standby" or "Measure".
1
Select the cross-hair icon
.
2
Select the calibration “Internal”.
3
Press
.
ð
The device performs an internal calibration.
6.3.16.2 External calibration
ü
Operator or Supervisor rights
Summary of Contents for ULTRATEST UL3000 Fab
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 133: ...INFICON Appendix 11 UL3000 Fab Operating instructions iina73en1 08 1910 133 140 ...
Page 134: ...11 Appendix INFICON 134 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 138: ...Index INFICON 138 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 139: ......
Page 140: ......