INFICON
Description | 4
UL3000-Fab-Operating-instructions-iina73en1-08-(1910)
31 / 140
UL3000 Fab, UL3000 Fab PLUS
Minimum detectable leak rate sniffer
mode
< 5 x 10
-8
mbar l/s
Gas flow through sniffer line SL200
25 sccm
Gas flow through sniffer line SL3000
160 sccm
Measurement range
12 decades
Sound pressure level in Standby
< 61 dB (A) *)
Run-up time
< 3 min
Ion source
2 long-life iridium cathodes, Yttrium oxide
coated
*) The A-weighted emission sound pressure level at the operator's premises is less
than 61 dB (A) at all times for all foreseeable uses of the device. The noise emission
measurement declaration was drawn up in accordance with the harmonized standard
DIN EN ISO 3744:2011.
Ambient conditions
UL3000 Fab, UL3000 Fab PLUS
Max. altitude above sea level
2000 m
Maximum relative humidity
80% at 30°C, linear decreasing to 50% at
40°C
Storage temperature
0°C to +50°C
Operating temperature
+10°C to +40°C
Pollution degree
2
4.5 Factory settings
Leak rate setpoint 1
1 x10
-9
mbar l/s
Leak rate setpoint 2
1 x10
-8
mbar l/s
Leak rate setpoint 3
1 x10
-7
mbar l/s
Leak rate setpoint 4
1 x10
-6
mbar l/s
Automatic purge
On
Volume
2
ZERO mode
I•ZERO 2.0
Calibration request
Off
TMP ventilation
On (can only be changed by service)
Interfaces leak rate unit vacuum operation
mbar l/s
Interfaces leak rate unit sniffer operation
mbar l/s
Interface pressure unit
mbar
Summary of Contents for ULTRATEST UL3000 Fab
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 133: ...INFICON Appendix 11 UL3000 Fab Operating instructions iina73en1 08 1910 133 140 ...
Page 134: ...11 Appendix INFICON 134 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 138: ...Index INFICON 138 140 UL3000 Fab Operating instructions iina73en1 08 1910 ...
Page 139: ......
Page 140: ......