Maintenance Work
103
iina
70
e 0
9
.f
m
te
chn
ica
l ha
nd
boo
k
(1
40
8)
Approximately 2.5 hours will be needed for this maintenance work, but will vary in
case that operating agent reservoir is exchanged.
9.5.3
8000 Hours Maintenance
The 8000 hours maintenance work should be done by an INFICON service
technician or a person authorised by INFICON.
For the scroll module of the Agilent scroll pump, the „tip seal“ must be replaced by
an INFICON service technician as soon as 8000 operating hours are attained. If the
„tip seal“ is not replaced, it will become necessary to replace the scroll module after
12000 operating hours.
Approximately 6.0 hours will be needed for this maintenance work. For the
replacement of the „Tip Seal“ additionally 3 hours are necessary.
9.5.4
16000 Hours Maintenance
The 16000 hours maintenance work should be done by an INFICON service
technician or a person authorised by INFICON.
As soon as 16000 operating hours have been attained, the life expectancy of the
various backing pump types has been reached.
Work to be done
Required materials
Part No.
Exchange the oil of the D16 backing
pump (UL1000 only)
Arctic Oil 1l
20099091
Check/replace the filter
Replacement filter cell for fan
Replacement filter for venting and
purge gas line
Silencer for exhaust (UL1000 Fab
only)
Oil filter insert (10 pcs.) UL1000 only
200000685
200000683
20099183
200000694
Test and alignment
Replace operating agent reservoir of
SplitFlow 80 only in case a 3 year
time period has elapsed. Please
refer to
.
Operating agent reservoir for
SplitFlow 80
200003801
Work to be done
Required materials
Part No.
Replace tip seal (Agilent TS 620)
Tip seal
200001671
Exchange oil for the D16 backing
pump (UL1000)
Arctic Oil 1l
20099091
Replace valve seals
Seal kit for valves
200000594
Check/replace filter
Replacement filter cell for fan
Replacement filter for venting and
purge gas line
Silencer for exhaust (UL1000 Fab
only)
Oil filter insert (10 pcs.) UL1000 only
200000685
200000683
20099183
200000694
Test and alignment
Replace operating agent reservoir of
SplitFlow 80 only in case if a 3 year
time period has elapsed. Please
refer to
.
Operating agent reservoir for
SplitFlow 80
200003801