2 – 2
Transpector SPS Operating Manual
2
Note the recessed area on the sensor feedthrough and the ground tab on the
Transpector electronics module. Match the recessed area of the feedthrough
to the ground tab and carefully slide the Transpector electronics module fully
onto the sensor.
3
Finger tighten the mounting nut on the Transpector sensor.
Figure 2-1 Transpector electronics module dimensions
2.2.2 Attach Heat Guard
To decrease the risk of burns when the heating jacket is on, a heat guard is
provided for the metal surfaces between the Transpector electronics box and the
Transpector SPS manifold tee.
1
Loosen the single screw that attaches the calibration reference shipping
bracket to remove the bracket. Replace the screw.
2
Attach the heat guard between the Transpector electronics module and the
mounting nut using the two screws provided.
3
Attach the calibration reference standard bracket to the heat guard and secure
the bracket to the heat guard with the provided screw.
Mounting Nut
SPS Manifold Tee
Transpector Electronics Module
Cable Box
Transpector Sensor