5 - 20
Transpector MPS Operating Manual
5.5.2.3.2 Default Tune Mode Parameters
When first entering the
TUNE
window, Transpector Web UI is set with the default
parameters displayed in
.
5.5.2.3.3 Tuning Transpector MPS
During normal operation, the tune should not deviate much from the original factory
tune. This tuning guide will only discuss fine tuning of the sensor. If the tune has
been altered significantly, contact INFICON.
In general, tuning Transpector MPS involves mass alignment (ensuring that the
peak is aligned with the proper AMU) and width adjustment.
Tuning RGAs is not something that should be done by novice users. For general
tuning instructions, see
. These tips and instructions
should be reviewed prior to doing any tuning regardless of the interface.
Instructions for tuning Transpector MPS using the Transpector Web UI interface
are as follows:
1
Click
CHANGE TUNE MASSES
. A tune table will display, see
.
To remove masses from the tune table, click the mass to remove, then click
REMOVE
.
CAUTION
Do not remove masses 1, 2 or 4. Removing the low tune
masses can significantly alter the higher-mass tuning.
To add masses to the tune table, type the mass into the box next to
ADD
,
then click
ADD
. The default Tune masses are displayed in
.
Using the tune table, select the masses that will be tuned.
Table 5-4 Default Tune Mode parameters
Parameter
Value
Dwell
32 ms
ppAMU
25
Electron Energy
70 eV
Scan Mass
1 AMU
Scan Width
±2 AMU