INFICON
Cleaning and maintenance | 8
Pernicka-700H-CHLD-Operating-instructions-jin76aen1-04-(1911)
85 / 100
2
Return the calibration leak to the manufacturer for recalibration.
3
Re-install the recalibrated calibration leak into the device. Please contact the
INFICON service.
8.5 Maintenance of the backing pump
8.5.1 Check the oil level
Carry out the maintenance work in accordance with the maintenance plan (see
87]") and the manufacturer's technical documentation delivered
with the device. See "Operating Instructions GA01601_002_A1".
Required spare parts
None
Required tools
None
The oil level of the backing pump must always be within the "minimum oil level" and
"maximum oil level" markings during operation.
► Check that the oil level is approximately in the center of the sight glass when the
backing pump is switched off.
Oil level mark
maximum
Oil level mark
minimum
Backing pump
Sight glass
Fig. 56: Backing pump: Check the oil level