Cleaning and maintenance | 8
Pernicka-700H-CHLD-Operating-instructions-jin76aen1-04-(1911)
79 / 100
Required tools
Aids for levering the O-rings, e.g. plastic tweezers or forceps
1
Open the lid of the test chamber.
2
Carefully remove the two O-rings from the test chamber using a suitable aid. Lift
the O-rings only upwards.
3
Clean the surface of the test chamber with a dust- and lint-free cloth.
4
Insert the two new O-rings into the test chamber.
5
Close the lid of the test chamber.
Test chamber
O-ring
Fig. 49: Test chamber: Lid open
8.3 Maintenance of the CHLD sensor
8.3.1 Check the fan function
Read and follow the included technical documentation "Operation Manual CHLD
Sensor QME 220".
The fan is used to cool the internal electronic modules of the CHLD-Sensor.
Required spare parts
See "Check or change the filter mat
81]"
See "Check or change the filter mat
81]"
Required tools
See "Check or change the filter mat [
► Check the fan function in the following situations: