11 | Troubleshooting
INFICON
298 / 319
074-594-P1H Micro GC Fusion Operating Man-
ual
2 Adjust column head pressure so that all peaks elute within method defined peak
windows.
3 Save the method.
11.2.2.2 Correcting Instrument Parameter Settings - Recalibrating
Changes to Column Pressure, Column Temperature, Inject (ms) time, and Backflush
Time (if applicable) may cause peak retention times to shift and peak areas to change.
The retention times and associated peak windows will require adjustment. Therefore,
the method should be recalibrated. Recalibrate Micro GC Fusion as instructed in
Calibration [
200].