44
tina83e1-d (2018-12)
Using an ohmmeter, make following measurements on the
contact pins.
Measurement
between pins
Possible cause
1 + 4
39.5 … 40.5
Ω
(at 20 °C)
Values
outside of
the range
Pirani filament
rupture
1 + 2
1000 … 1100
Ω
(at 20 °C)
Values
outside of
the range
Pirani tempera-
ture sensor
rupture
5 + measuring
chamber
∞
<<
∞
Contamination,
short circuit cold
cathode
1
2
3
4
5
Measuring chamber
All of theses faults can only be remedied by replacing the
complete measuring chamber (
→
We recommend to perform a leak test (leak rate
<1×10
-9
mbar l/s).
6.2.2
Replacing Ionization Chamber and Ignition Aid
Precondition
Troubleshooting (measuring chamber) performed (
→
Summary of Contents for Gemini MAG500