6.2.7 Switching automatic measurement start on or off
The option "Autostart" is activated in the factory settings. If you select the function
"Measurement" and then close the measuring chamber, the selected process is
started automatically. The signal from a proximity switch is used for this purpose. You
can switch the "Autostart" option on or off.
ü
Operator or Supervisor rights
1
> Device
2
Adjust.
3
Save
.
ð
If the automatic measurement start is OFF, press the "START" button on the
touchscreen or on the housing to start the measurement.
6.3 Settings for the measurements
6.3.1 Select product
1
> Products
ð
Existing products are displayed. If you cannot find the desired product, you
can create it, see "Select, modify or create product (measurement settings)".
Click the desired product name.
2
Load
.
6.3.2 Perform ZERO measurement
Use this function for low levels of contamination. The current background value is then
set to zero. If there are high background values, first use the "Purge" function.
The result is displayed in green and the values are applied. A warning or an error is
displayed in the event of a problem.
1
Select
>
.
2
Empty the measuring chamber.
3
Start the ZERO measurement.
ð
The result is displayed in green and the values are applied. In the event of an
error, the result is displayed in red.