6.3.4 Check the Operation of the Device
You have the possibility to check the function of the device with the help of the
supplied CON-Check (test specimen).
The CON-Check simulates leak rates in the range of 0.016 ± 0.004 mbar l/s.
If you want to check with a higher leak rate, you have the alternative option to place
multiple CON-Checks or a Pac-Check in the measuring chamber. In order to ensure
the proper functioning of the device (measurement result: Leaky, red), adjust the
maximum leak rate accordingly.
1
> Products
ð
In addition to other products, the product "CON-Check" will be shown
according to factory settings.
2
Click the product name "CON-Check".
3
Load
.
ð
You will be asked to do a ZERO measurement with an empty measuring
chamber, see "Perform ZERO measurement [
ð
Loading the Con-Check sets a permissible maximum leak rate that is less
than the leak rate simulated by the Con-Check. This allows the result of the
test to report a leak.
4
Place the CON-Check in the measuring chamber on its own.
5
Start the measurement process.
ð
If the CON-Check is identified as leaking for exceeding the maximum permissible
leak rate, the device is working properly, see "Result display [
18]”.
The measured leak rate should be in the range of 0.016 ± 0.004 mbar l/s.