background image

1 of 8

CMS5000 Monitoring System Concentrator
Installation Instructions

INSTRUCTION SHEET

PN 074-536-P1C

CAUTION

The CMS5000 concentrator is not interchangeable with the 
concentrators from the HAPSITE

®

 family of products.

1

Disconnect the power by unplugging the Fischer Connector

®

 from the right side of 

CMS5000. Open the front panel by unlocking (if necessary) and rotating the latch 
clockwise. See 

Figure 1

.

NOTE:

Nitrile or equivalent gloves are recommended when replacing the 
concentrator.

Figure 1  Disconnect power, open the front panel

Summary of Contents for CMS5000

Page 1: ... the concentrators from the HAPSITE family of products 1 Disconnect the power by unplugging the Fischer Connector from the right side of CMS5000 Open the front panel by unlocking if necessary and rotating the latch clockwise See Figure 1 NOTE Nitrile or equivalent gloves are recommended when replacing the concentrator Figure 1 Disconnect power open the front panel ...

Page 2: ...em Concentrator Installation Instructions 2 Using a screwdriver remove the three screws that hold the lower oven Retain the screws for reuse See Figure 2 Remove the lower oven cover See Figure 3 Figure 2 Open the front panel Figure 3 Remove cover ...

Page 3: ...crews that hold the upper oven cover in place Retain the screws for reuse Remove the upper oven cover See Figure 4 Figure 4 Remove four screws and upper oven cover 4 Rotate so the board is parallel to the unit gently grasp the board with both fingers and pull straight out See Figure 5 Figure 5 Remove concentrator board ...

Page 4: ...the right side upwards and the nut on the left side downwards Using fingers fully loosen both ends of the concentrator See Figure 6 Figure 6 Loosen the nuts 6 Using a 3 8 wrench stabilize the concentrator elbow on the right side Using a screwdriver unscrew and remove the two screws that hold the elbow in place Retain the screws for use when reinstalling See Figure 7 Figure 7 Remove two screws ...

Page 5: ...elbows See Figure 8 Figure 8 Remove from the concentrator elbows CAUTION The concentrator is directional The end with the groove in the metal band must face toward the right when installed There are two small ferrules that are seated within the nuts The tapered end of the ferrules must face outward See Figure 9 Figure 9 Concentrator direction ...

Page 6: ...right concentrator elbow place the grooved end of the concentrator into it C See Figure 10 Figure 10 Replace concentrator 9 While holding the right concentrator elbow in place with your fingers insert and tighten the bottom screw until finger tight Continue to stabilize the right concentrator elbow in place with a 3 8 wrench insert the top screw and evenly tighten the screws until snug Do not over...

Page 7: ...n both sides Using a 7 16 wrench tighten both nuts by 1 4 of a turn Tighten the nut on the right side downwards and the nut of the left side upwards See Figure 12 Figure 12 Ensure that the concentrator is fully seated and tighten nuts 11 Line up the clips from the concentrator board with the metal contacts on the concentrator See Figure 13 Figure 13 Line up the clips ...

Page 8: ...ns 12 Press the concentrator board gently but firmly onto the concentrator See Figure 14 Figure 14 Press the concentrator board 13 Position the upper oven cover as shown in Figure 15 Using a screwdriver screw the four screws into the proper locations Position the lower oven cover as shown Screw the three screws into the proper locations Figure 15 Replace the upper and lower oven covers 14 Shut the...