6 | Operation
INFICON
52 / 128
IRwin-Operating-instructions-mina66en1-15-(1904)
• The gas must be delivered at ambient pressure (i.e., do not push gas into the leak
detector). The normal way to accomplish this is to create an open connection
between gas container and detector inlet so that the excess calibration gas is
vented to the ambient. This exhaust should be placed outside the test room,
preferably outdoors.
Calibration
ü
Login as Intermediate or Advanced
►
>
> Calibration > Calibration (Tab)
ð
Select the calibration to perform and click the check mark button.
ð
The leak detector leads you through the process by indicating which gas to
deliver to the detector in each step. Make sure that no residual gases are
present in the ambient during Fresh Air step. Most common problems occur if
you used 2.2 Vol% or 100 Vol% methane shortly before the trying to calibrate
with 10 ppm or 100 ppm methane.
ð
A successful calibration is confirmed by a green check mark.
ð
A failed calibration is indicated by a red cross.
You can continue to use the leak detector even if calibration failed. The previous
calibration parameters are restored.
6.2.15.2 Setting of calibration gas concentrations
Accuracy can be affected by calibration gas tolerance.
For optimum performance adjust the concentrations of the calibration gases set in the
leak detector to the actual composition of your calibration gases.
For optimum performance of your leak detector you should use the proposed
calibration gases. The exact concentrations of the gas delivered often deviates slightly
from these values due to tolerance factors etc. The exact value is typically stated on
the analysis certificate of the respective gas. You should therefore adjust the
concentrations of the calibration gases set in the detector to match the actual
composition of your calibration gases, see "Concentration of calibration and test gases
ü
Login as Intermediate or Advanced
1
>
> Calibration > Settings (Tab)
2
Select calibration routine first and then select the gas to adjust (if several).
3
Confirm your selection with
.
ð
A window appears with controls for adjusting the gas selected.
Summary of Contents for 580-000
Page 2: ...INFICON GmbH Bonner Strasse 498 50968 Cologne Germany ...
Page 99: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 99 128 ...
Page 101: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 101 128 ...
Page 102: ...9 Certificates INFICON 102 128 IRwin Operating instructions mina66en1 15 1904 ...
Page 103: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 103 128 ...
Page 105: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 105 128 ...
Page 110: ...9 Certificates INFICON 110 128 IRwin Operating instructions mina66en1 15 1904 ...
Page 111: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 111 128 ...
Page 113: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 113 128 ...
Page 114: ...9 Certificates INFICON 114 128 IRwin Operating instructions mina66en1 15 1904 ...
Page 115: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 115 128 ...
Page 116: ...9 Certificates INFICON 116 128 IRwin Operating instructions mina66en1 15 1904 ...
Page 117: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 117 128 ...
Page 119: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 119 128 ...
Page 120: ...9 Certificates INFICON 120 128 IRwin Operating instructions mina66en1 15 1904 ...
Page 121: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 121 128 ...
Page 122: ...9 Certificates INFICON 122 128 IRwin Operating instructions mina66en1 15 1904 ...
Page 123: ...INFICON Certificates 9 IRwin Operating instructions mina66en1 15 1904 123 128 ...
Page 126: ...Index INFICON 126 128 IRwin Operating instructions mina66en1 15 1904 ...
Page 127: ......
Page 128: ......