INFICON
Operation | 6
T-Guard2_Operating-instructions-jina85en1-13-(2009)
47 / 90
6.5 Operation modes
You can carry out measurements in 3 operating modes:
• Accumulation measurement
• Carrier gas measurement
• Continuous measurement (accumulation or carrier gas)
6.5.1 Accumulation measurement
If a leaking test object that has been filled with helium under pressure is placed in a
closed measurement chamber, the helium concentration in this measurement
chamber increases over time. This is called accumulation (of helium). The difference
between the helium concentrations at 2 time points provides the leak rate.
The air in the measurement chamber must be circulated so that the concentration of
the helium in the entire measurement chamber is set equally.
The following diagram shows the typical helium concentration over time in a closed
measurement chamber with a leaking test object. Because of the helium content in the
environment, the helium value is never zero. The helium concentration shows linear
growth over time. The device measures the helium concentration in the measurement
chamber 4 times (except for turbo measurement).