TRIGGER
Through this menu item the following settings can be entered:
LEVEL
(rejection leak rate)
ALT
LEVEL
GAS
TYPE
(
with Smart probe only
)
BUTTON
UNIT
These settings are described in the following:
LEVEL (rejection leak rate)
The "
LEVEL
” menu item is selected when wanting to enter the leak rate at which the
units under test shall be rejected. The level may be set 1 g/a (0.04 oz/yr) (depending
on type of gas) and 50 g/a (1.76 oz/yr). For further units please refer to the following
table:
(Values in parenthesis are for Smart probe only)
ALT LEVEL
Entry of the alternative rejection rate. This rejection rate is used only when the key
on the handle has been programmed to "
ALT
” (see below). This setting may be
entered representing a percentage of the value for
LEVEL
, for example 50 %, or in
absolute terms, for example 10 g/a.
GAS TYPE (For Smart
probe only)
The HLD5000 with Smart probe offers 5 pre-programmed gases to choose from.
When opening the gas type menu a list of these five gases plus three user-definable
gases is displayed. The bottom two buttons on the left side of the display allow to
scroll the list. The currently selected gas is highlighted by inverted colors.
When switching between the pre-programmed gases the HLD5000 does not require
to be recalibrated. The HLD5000 will automatically adjust for the different sensitivities
the Smart probe has for the different gases. However, the trigger value will stay the
same.
When choosing a user definable gas the HLD5000 can either be calibrated externally
or a calibration factor can be entered to enable a calibration with the built-in
calibrated leak. These factors have to be determined for each additional gas and may
be obtained from INFICON on request in most cases. When not entering a calibration
Unit
Lower display limit
Upper display limit
g/a
(0.5) 1.0
50.0
mbar l/s
-1
4 x 10
-6
3.9 x 10
-4
lb/yr
(1) 2 x 10
-3
1.0 x 10
-1
oz/yr
(0.02) 0.04
1.76
Pa m
3
x s
-1
4 x 10
-7
3.9 x 10
-5
Summary of Contents for 510-010
Page 2: ...0 2 kina40e IVZ fm technical handbook 1007 ...
Page 8: ...0 8 General Safety Precautions kina40e IVZ fm technical handbook 1007 ...
Page 40: ...4 12 HLD5000 Settings Menu Structure kina40e 04 fm technical handbook 1007 ...
Page 48: ...6 6 Maintenance kina40e 06 fm technical handbook 1007 Fig 6 2 Declaration of Conformity ...
Page 51: ...Appendix A 3 kina40eSIX fm technical handbook 1007 ...