Technical Data
1.2.1
Physical Data
1.2.2
Electrical Data
1.2.3
Other Data
Lowest trigger level setting
For single gas probes
1.0 g/a (0.04 oz/yr)
For universal Smart probe
0.5 g/a (0.02 oz/yr)
Maximum trigger level setting
50 g/a (1.76 oz/yr)
Measurement range
For single gas probes
0 - 100 g/a (3.57 oz/yr)
For universal Smart probe
0 - 300 g/a (10.7 oz/yr)
Detectable refrigerants
Single gas probe
(depending on the instrument version)
R134a,
R744 (CO
2
), SF
6
Universal Smart probe
all halogens
Response time
1 s
Time until ready for operation
30 s
Recovery time
approx. 2 s
Gross leak recovery time
approx. 8 s
Mains supply voltage
100 V ... 230 V ± 10 %
50/60 Hz
Power consumption
in standby
< 60 VA
< 40 VA
Protection
IP 30
Overvoltage category
II
Gas flow at the intake
320 sccm
Dimensions (height, diameter)
365 mm; 260 mm
(14.4 in.; 10.25 in.)
Weight of the main unit
4.5 kg (10 lb.)
Weight of the probe
390 g (0.9 lb.)
Permissible ambient temperature range
(during operation)
5 – 50 °C; 40 – 120 °F
Max. rel. humidity of the air
80 % at +31 °C, decreasing
to 50% at 40°C (104°F)
Length of the sniffer line
4.8 m (15.5 ft.)
Summary of Contents for 510-010
Page 2: ...0 2 kina40e IVZ fm technical handbook 1007 ...
Page 8: ...0 8 General Safety Precautions kina40e IVZ fm technical handbook 1007 ...
Page 40: ...4 12 HLD5000 Settings Menu Structure kina40e 04 fm technical handbook 1007 ...
Page 48: ...6 6 Maintenance kina40e 06 fm technical handbook 1007 Fig 6 2 Declaration of Conformity ...
Page 51: ...Appendix A 3 kina40eSIX fm technical handbook 1007 ...