50
tinb60e1 (2020-07)
7.5
Allocating DeviceNet connections to the
PGE500
The master must use a DeviceNet object command to open
Polled and Explicit connections. Other connection types are not
supported. The example below allocates polled and explicit
connections.
Service
4B hex
DeviceNet Class
3
Instance
1
Allocation data
Bit1=Polled, bit0=Explicit
3
Master ID
0
7.6
Configuring Expected Packet Rate
The default Expected Packet Rate is 2.5 seconds. If message
rate is slower this attribute must be changed before the allocated
connections expire. The example below sets the Explicit packet
rate to never expire. Use instance 2 to change the Polled packet
rate.
Service
10 hex
Connection Class
3
Instance
1
Attribute
9
Data
0