10
Install 1/8 in. sample input line to the top right Swagelok fitting on the accessory.
See
.
Figure 5-13 Installing sample input and vent lines
WARNING
If a hazardous sample is used, the sample exhaust must
be vented to a fume hood or other designated area for
hazardous waste disposal.
11
Connect a vent line to the vent fitting using standard 1/8 in. fittings. Run the
exhaust to a fume hood or other designated area for hazardous waste disposal.
Installation is now complete.
shows the Pressure Reducer mounted on 3000 Micro GC.
Figure 5-14 Pressure reducer mounted on a 3000 Micro GC
Install sample input line here
Install vent line here