3000 Micro GC Operating Manual
Chapter 5
Accessory Installation
5.1 Sample Conditioners
3000 Micro GC can be damaged by contaminants, especially particulates and
condensing aerosols. Use an appropriate filter or sample conditioner at all times.
lists the typical usage and filtration capabilities of the standard filter and
available accessories.
5.1.1 List of Sample Conditioners and IPNs
Table 5-1 List of sample conditioners, inlet pressures, and IPNs
Type
Input Pressure
to Accessory
Sample
Container/delivery
Sample Matrix
Particle
filtration
(microns)
PN
Standard external
sample inlet filter, or
direct connection
0 to 172 kPa
(0 to 25 PSI)
All
Relatively clean
and dry
10
G2801-60900
Genie Filter
Assembly
70 to 345 kPa
(10 to 50 PSI)
All
Entrained liquids
and particles
-
G2817A
Pressure Reducer
345 to 6895 kPa
(Ambient to
1000 PSI)
All
C6+ components
<0.5 mole%
-
G2815A
Pressure Reducer
and Genie Filter
Assembly
<3477 kPa
(<500 PSI)
All
Entrained liquids
and particles,
C6+ components
<0.5 mole%
G2816A
Summary of Contents for 3000 Micro GC
Page 1: ...O P E R A T I N G M A N U A L 3000 Micro GC Gas Analyzer PN 074 519 P1C Cover Page ...
Page 2: ......
Page 6: ......
Page 8: ......
Page 18: ...TOC 10 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Page 26: ...1 8 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Page 128: ...6 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...
Page 268: ...11 20 PN 074 519 P1C 3000 Micro GC Operating Manual This page is intentionally blank ...