Industrial Test Systems 486691 Instruction Manual Download Page 4

4

DPD-1 Free Chlorine Test Procedure

1

6a

6b

5

2a

3

4

DO NOT

 discard the sample from the Free Chlorine test,

if you are planning to run eXact

®

 Strip Micro DPD-3 (Total

Chlorine) Procedure, Continue steps 7-9 on page 5.
Otherwise, rinse the cell immediately.

2b

CL1:   Chlorine reacts with N,N-diethyl-p-phenylenediamine 
as it is released from the strip to form a magenta color, 
directly proportional to the Chlorine concentration. (Ozone, 
Total Chlorine, and Permanganate are also determined in 
the CL1 MENU)

Remove one (1) 

eXact

®

 Strip Micro CL (DPD-1)

, part

number 486637, from the bottle before beginning the
test. Set the strip in a dry, convenient place and recap
the bottle immediately.

REMOVE STRIP

1

Press the 

ZERO/ON

 button to power the meter on; the

display will show all annunciators, then the current
MENU selection, followed by the last reading.

TURN METER ON

2

Dip the 

eXact

® 

Strip Micro CL (DPD-1), Part No.

486637

 into the 

CELL

 and  immediately press 

READ

.

This starts the 

20 SECOND

 countdown timer. During

this time move the strip in a gentle back and forth
motion. 

Remove and discard the strip when the time

displays 1

. The cursor will move across the display,

informing you that it is about to measure the sample.
Record result displayed (this result is automatically
stored in CL1).

DIP STRIP AND PRESS “READ”

6

Press the 

ZERO/ON

 button. The cursor will move

across the display followed by 

0.00 PPM

. Sample is

ready for testing.

ZERO METER

5

Rinse the 

CELL

 at least 3 times with the water sample

you will be testing - rinsing minimizes the potential for
cross-contamination from a previous test. Finally, fill
cell to capacity (4ml) with the water sample.

FILL METER WITH SAMPLE

4

Press and re-press the 

MENU

 button until the

display shows the parameter you want to run.

CL1 

is also used for testing:

Total Chlorine (DPD-4), Ozone (DPD-4), Permanganate (DPD-1), and
Total Chlorine (DPD-3). (Contact ITS for specs and details if you are
planning on using 

CL1 

for Permanganate or Ozone measurements)

SELECT TEST: CL1

3

CL

CL1

29

1.

2.

3.

4.

5.

6.

7.

8.

9.

10.

11.

12.

13.

14.

15.

16.

17.

18.

19.

20.

21.

22.

Become familiar with the meter and the different tests by reading the instructions carefully.

The DPD-1 & DPD-4 strips supply enough reagent for accurate results at room temperature. If 
samples are below 60°F (15°C) the 20 second strip dip time will give low results for chlorine 
values above 4 ppm. For cold samples, best results are obtained when a 40 second dip time is 
used (Dip one strip for two (2) twenty second periods).

The Free Chlorine and Total Chlorine results are compliant for meeting USEPA (4500-Cl G); ISO 
7393/2; and German DIN 38408 G4-2 requirements.

Observe the dip time (

as required for the test

) for accurate results.

Test immediately after filling the 

CELL 

with water sample when testing for an oxidizer such as 

Chlorine, Bromine, Ozone, etc...

Be sure the 

CELL 

is filled to capacity (4ml), especially for pH and Total Alkalinity.

Sample water that may splash out of the 

CELL

 during movement of the eXact

®

 Strip Micro will not 

affect accuracy, as long as the cell is 50% full at end of test.

Rinse the 

CELL 

with clean water immediately after completing the test. 

(Some test reagents will 

stain or coat the CELL)

Just before testing, rinse the sample 

CELL 

with the sample water several times to get a 

representative sample. (

Use deionized or distilled water for rinsing if you have a limited amount of 

sample

).

Store the meter and all test materials out of direct sunlight and away from chemical storage areas.

Minimize exposure of meter and test reagents to heat above 100°F (38°C).

Dry the outside of the meter when testing is complete or before storage of the meter.

When running a DPD-1 Free Chlorine test 

AFTER

 a Total Chlorine DPD-3 or DPD-4 test, rinsing is 

very important to remove residual KI, which may interfere.

Each eXact

®

 Strip Micro is valid for 

ONLY

 one test. Discard strip after single use in regular trash 

that is inaccessible to children and pets.

Each bottle of eXact

®

 Strip Micro contains the quantity of strips notated on the bottle. Due the 

manufacturing process, you may find one or two strips that are noticeably smaller or larger in 
width than the normal strips in the bottle. These should be discarded. Using these strips may give 
unreliable results.

Each conversion table supplied has a unique revision number located in the bottom right corner of 
the chart. It is recommended that you visit www.sensafe.com at least every 6 months to check for 
any updated revisions.

Tests are calibrated at 75°F +/- 2°F (24°C +/- 1°C). If water sample is 60°F (15°C) or cooler, most 
tests like Copper, Manganese, Nitrate, and Nitrite may require additional time for full chemical 
development. At the end of the normal test procedure, press 

READ

 again and compare value after 

this countdown. If the new value is higher, then use the new value for your result. It is 
recommended that the water sample be warmed to room temperature before testing.

The eXact

®

 Micro 7+ Meter is not compatible for use with DPD-1, DPD-3, and DPD-4 powder 

pillows, tablets, and liquids available from other manufacturers. Accurate results can only be 
guaranteed by using genuine eXact

®

 Micro strips or reagents (reorder information on page 31).

Normally the mixing cap is used for tests like Cyanuric Acid, Fluoride, Iron, and Manganese. A 
stirrer (such as a clean, plastic coffee stirrer) may also be used in place of the mixing cap by using 
the stirrer with gentle, back and forth motion, during the 20-second countdown timing.

Our lab testing with the Micro 7+ meter has shown that zeroing and measuring of the sample does 
not require any cell cover for accurate results, even in full sunlight.

Remove batteries when meter is not used for more than a month.

Pool and Spa samples should be taken 18 inches below the surface as follows; submerge meter 
with open cell facing down 18 inches, and then turn meter upright to fill cell. Remove meter from 
water with the sample for testing.

eXact

®

 Micro 7+ Tips For Best Accuracy

Summary of Contents for 486691

Page 1: ...3 329 9743 ITS SENSAFE COM WWW SENSAFE COM www poolcheckonline com For European Middle East Inquiries and Re Orders ITS Europe LTD Unit B Stanley Crt Glenmore Business Park Telford Rd Churchfields Industrial Estate Salisbury Wiltshire SP2 7GL United Kingdom Tel 44 0 1722 329502 Fax 44 0 1722 329880 SALES SENSAFE COM WWW ITSEUROPE CO UK www poolcheckonline com eXact Strip Micro DPD 1 Accuracy eXact...

Page 2: ...7 to be run first Alizarin Red S Citrate DPD Oxalic Acid DPD Kl DPD Kl Biquinoline DPD Kl DPD Phenol Red Salicylate Method Silver ppt DPD Diphenylcarbazide Melamine ppt SPADNS DPD TPTZ 1 10 Phenanthroline Phthalein Purple PAN Cyanide Zinc Reduction Chromotropic Acid Alizarin Red S Thymol Blue Tetraphenylborate Barium ppt Turbidity values as NTU PARAMETER No PART NUMBER OF TESTS DETECTION RANGE CHE...

Page 3: ...0 The following are some common messages that may be displayed including error messages If an error message other than those listed below is displayed please contact technical support in the USA at 803 329 0162 ext 0 LCD Message HI LO LO ER Description In READ mode test sample concentration is above the measurement range test specific In READ mode test sample concentration is below the measurement...

Page 4: ... for pH and Total Alkalinity Sample water that may splash out of the CELL during movement of the eXact Strip Micro will not affect accuracy as long as the cell is 50 full at end of test Rinse the CELL with clean water immediately after completing the test Some test reagents will stain or coat the CELL Just before testing rinse the sample CELL with the sample water several times to get a representa...

Page 5: ... time the KI is added For compliance testing you must time the two minutes and then make your measurement NOTE From testing in our lab water samples above 70 F 20 C generally reach a stabilized reading quickly eXact Strip Micro CL DPD 1 DPD 3 DPD 4 Interferences part nos 486637 486638 486670 Interfering Substance Acidity Alkalinity Bromine Bromamines Br2 Chlorine Dioxide ClO2 Copper Cu 2 Iodine I2...

Page 6: ...y 0 01 press READ again and repeat After testing is completed rinse cell immediately Record the Total Chlorine as the highest value the meter displayed NOTE Standard Method 4500 Cl G procedure for total chlorine requires the reading to be made 2 minutes after the reagents are added pH Test Procedure 1 Remove one 1 eXact Strip Micro PH part number 486639 from the bottle before beginning the test Se...

Page 7: ...n the display will show all annunciators then the current MENU selection followed by the last reading TURN METER ON 3 Press and re press the MENU button until the display shows the parameter AL4 SELECT TEST AL4 4 Rinse the CELL at least 3 times with the water sample you will be testing rinsing minimizes the potential for cross contamination from a previous test Finally fill cell to capacity 4ml wi...

Page 8: ...TER WITH SAMPLE Copper Test Procedure CA5 Calcium reacts with Oxalic acid as it is released from the strip to form a white precipitate directly proportional to the Calcium concentration Values are reported as Calcium Carbonate CU6 Copper reacts with Biquinoline or Bicinchoninic Acid as it is released from the strip to form a purple color directly proportional to the copper concentration CA CA5 CU ...

Page 9: ... charts round off your T value to two digits Tests that can be used with the Micro 7 in TR7 MENU are listed on page 31 Typically the most common tests are supplied with the conversion chart in this booklet For conversion charts and information about the tests not given in this booklet visit our website You can also use the TR7 MENU to develop your own custom conversion chart for unusual samples to...

Page 10: ...TER WITH DISTILLED OR DEIONIZED WATER Turbidity Test Procedure TU TR7 Alkali pH Table Alkali pH results require the table below Follow eXact Micro 7 Alkali pH Test Procedure above using eXact Strip Micro Alkali pH Part No 486609 eXact Strip Micro Alkali pH Part No 486609 for 4mL Samples Rev 020509 AlkpH tr 90 80 70 60 50 40 30 20 10 0 9 8 7 6 5 4 3 2 1 0 8 3 8 3 8 3 8 3 8 5 8 7 8 9 9 1 9 3 9 6 8 3...

Page 11: ...14 18 23 29 39 60 5 5 8 11 15 19 24 30 40 60 5 5 8 12 15 19 24 31 42 60 5 6 9 12 15 20 25 31 45 60 Cyanuric Acid Test Procedure CY TR7 Iron as Fe 2 Table Iron results require the table below Follow eXact Micro 7 Iron as Fe 2 Test Procedure above using eXact Strip Micro FE Part No 486631 6 Find the TR result in the table below to determine the Iron concentration in ppm parts per million Example a T...

Page 12: ...0 03 0 03 0 39 1 06 1 84 2 8 4 1 5 8 5 8 5 8 12 NH3 TR7 Chlorine Dioxide Table Chlorine Dioxide results require the table below Follow eXact Micro 7 Chlorine Dioxide Test Procedure above using eXact Strip Micro CL DPD 1 Part No 486637 eXact Strip Micro CL DPD 1 Part No 486637 for 4mL Samples Rev 121908 ClO2 tr 90 80 70 60 50 40 30 20 10 0 9 8 7 6 5 4 3 2 1 0 0 03 0 23 0 51 0 82 1 19 1 62 2 2 2 9 4...

Page 13: ...hloride as NaCl Test Procedure For Tap Water CHLORIDE TR7 6 Dip the eXact Strip Micro CHLORIDE as NaCl Part No 481657 into the CELL and immediately press READ This starts the 20 SECOND countdown timer During this time move the strip in a gentle back and forth motion Remove and discard the strip when the time displays 1 The cursor will move across the display informing you that it is about to measu...

Page 14: ...ples with suspended solids and particulates may cause incorrect results since the reducing agent in the powder pillow often dissolves the suspended solids and particulates If the meter is zeroed with the original water sample the zero is no longer valid for this kind of sample So for this kind of sample zero the meter using a water sample free of suspended solids and particulates use distilled dei...

Page 15: ... 7 Find the TR result in the table below to determine the Manganese concentration in ppm parts per million Example a TR result of 65 3 use only the 65 for the chart equals a Manganese value of 0 04 ppm Record result After testing is completed rinse cell immediately USE TABLE 4 Press the ZERO ON button The cursor will move across the display followed by 100 T Sample is ready for testing ZERO METER ...

Page 16: ...8 39 62 80 100 121 140 140 140 140 8 41 64 82 102 124 140 140 140 140 8 43 66 84 104 126 140 140 140 140 13 46 68 86 106 128 140 140 140 140 18 49 70 88 108 131 140 140 140 140 NOTE Divide the above Nitrate result by 4 4 to determine Nitrate value as Nitrogen NO3 as N 16 NO3 TR7 Nitrite NO2 Table Nitrite results require the table below Follow eXact Micro 7 Nitrite NO2 Test Procedure above using eX...

Reviews: