Workplace requirements

 Location: Indoor use only. Any outdoor area where the equipment is to be

  

located should meet all the conditions below.

 Operation environment: 10~40

℃ 

(50~104

), 30~75%RH, 70~106kPa

 Adapter:  Power Input 100~240V, 50/60Hz, 1.2A 

                 Power Output DC 12V, 3.33A

Note on Unpacking & Assembling

(1) Place the BSM370 on level ground.
(2) Do not move by holding the head bar or operation part with LCD.

2. Raise the stand of the BSM370

 

Raise the stand of the BSM370 carefully until it stands upright. 

  Tighten the joint screw connecting stand body and foot plate.

4. Balance the BSM370

Use the bubble level indicator on the BSM370 to check it is level. Use the 

leveling screw underneath the foot plate of the BSM370 if it is not level.

 

Turn on the power switch. The head bar will move up automatically. 

  Head bar of BSM370 will be automatically went up after turning the

  device on.

3. Fixed the head bar

Unscrew the fixed screw at the head bar  

part and uplift it by 90 degree. 

Screw tightly.

1. Open the box

Open the packing box and remove the wrapping.

Take out all contents.

5. Power Connection

 

Connect the power adapter for the BSM370 which is placed on the

  back of the foot plate.

 

After the upper part of the BSM370 is fixed, 

  close the joint cover.

Bubble level indicator

Leveling screw

<Lowering>

<Raising>

<Un-leveled state>

<Leveled state>

screw

Remote control device is provided with thermal printer which is attached on.

NOTE

Save the wrapping material after unpacking in the event of relocation. 

NOTE

Do not lift the head bar to prevent any malfunctioning.

WARNING

Make sure you use the power adapter and main cable provided by 

InBody.

WARNING

If the equipment is not horizontally level, adjust the leveling screw on the 

opposite side of the bubble to place it at the center of the level indicator.

NOTE

In case of a loose connection between the stand body and the foot plate, 

measurement process cannot be performed properly. 

Fix the stand perpendicular on the foot plate.

CAUTION

Please check the height from the ground to the ceiling before installing 

BSM370. It should be at least as tall as 2156mm ( 84.9in.). 

BSM370’s height is 2126mm (83.7in.). 

CAUTION

When pressure or weight is applied to the foot plate during self-calibration,
              message appears. Turn the BSM370 off and on again after removing 
the material from the foot plate. Please do not apply pressure or weight to the 
foot plate during self-calibration.

Be careful not to put any weight on the foot plate. If you step on the foot plate 

or put a heavy object on the foot plate, it will affect the calibration and 

measurements will not be accurate.

CAUTION

NOTE

Quick Reference

Included Items

 

BSM370

 

Quick Reference

 

User’s manual CD

 

Adapter (12V, 3.33A) 1EA 

 

Power cord 

  (AC 250V, 10A, 1.8m( 5ft. 10in.)) 1EA

  

Screw

 

Joint cover

Optional 

 

Thermal printer 1EA , Thermal printer base

  Screw (M4

×

8) 6EA,  Cable (95cm (3ft. 1in.))

 

Remote control device, Cable (2m (6ft. 7in.),10cm (4in.))

 

Cable - Y type