34
35
E. EMC Declaration
Test Name
IEC 60601-
1-2
4th Edition
Ref. Standard Ports to Test AC Mains Volt-
age
Test Level Required
Notes
Mains terminal
disturbance voltage
4th
CISPR
11:2015/
AMD2:2019
AC Mains
100V, 50Hz
100V, 60Hz
120V, 60Hz
220V, 60Hz
230V, 50Hz
Group 1,
Class B
Radiated disturbance
4th
CISPR
11:2015/
AMD2:2019
Enclosure
100V, 50Hz
100V, 60Hz
120V, 60Hz
220V, 60Hz
230V, 50Hz
Battery
Group 1,
Class B
Harmonic Current
Emissions
4th
IEC 61000-3-
2:2018
AC Mains
230V~50Hz
Class A
Voltage Fluctuations
and Flicker
4th
IEC 61000-
3-3:2013/
AMD1:2017
AC Mains
230V~50Hz
Pst = 1
Plt = 0.65
Dmax = 4%
DC = 3.3%
Electrostatic
Discharges (ESD)
4th
IEC 61000-4-
2:2008
Enclosure
100V, 50Hz
100V, 60Hz
120V, 60Hz
220V, 60Hz
230V, 50Hz
Battery
±8 kv
contact
±2 kv, ±4 kv
±8 kv, ±15 kv
Air
1 pulse/
1sec
contact 8
kV
air 15kV
Radiated RF
elecromagnetic Fields
4th
IEC 61000-4-
2:2008
Enclosure
100V, 50Hz
100V, 60Hz
120V, 60Hz
220V, 60Hz
230V, 50Hz
Battery
10 V/m
80 MHz-2.7 GHz
80% AM at 1kHz
RF Wireless Comm.
(Refer to test report
clause 1.15)
Dwell Time
is 3 sec
Electric Fast
Transients and bursts
4th
IEC 61000-4-
4:2012
AC Mains
100V, 50Hz
100V, 60Hz
120V, 60Hz
220V, 60Hz
230V, 50Hz
±2 kv, AC,
100 hk PRR
5 pulses at
0°, 90°,
180°, 270°
Surges
4th
IEC 61000-
4-5:2014/
AMD1:2017
AC Mains
100V, 50Hz
100V, 60Hz
120V, 60Hz
220V, 60Hz
230V, 50Hz
±0.5 kv, ±1 kv
L to N (CM)
5 pulses at
0°, 90°,
180°, 270°
±0.5 kv, ±1 kv
±2 kv
L & N to PE (CM)
Conducted
Disturbances, induced
by RF fields
4th
IEC 61000-4-
6:2013
AC Mains &
all I/O
100V, 50Hz
100V, 60Hz
120V, 60Hz
220V, 60Hz
230V, 50Hz
Battery
3 Vrms
150 Hk to 80 Mk
80% AM at 1Hk,
2Hz
6 Vrms in ISM
Bands, amateur
radio
bands between
0.15 Mk and 80Mk
Dwell time
is
3 sec