42
Electromagnetic immunity
The [Model name] is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. Portable RF communications equipment should be used no closer than 30 cm
(12 inches) to any part of the [Model name]. Otherwise, degradation of the performance of this equipment
could result.
Immunity test
Band a)
Service a)
Modulation b)
IEC60601
test level
Compli
-
ance level
Proximity fields
from
RF wireless
Communica
-
tions
IEC61000-4-3
380 - 390 MHz TETRA 400
Pulse modula
-
tion b) 18Hz
27 V/m
27 V/m
430 – 470 MHz GMRS 460
FRS 460
FM c)
±5 kHz
deviation
1 kHz sine
28 V/m
28 V/m
704 – 787 MHz LTE Band13, 17
Pulse modula
-
tion b) 217 Hz
9 V/m
9 V/m
800 – 960 MHz GSM800:900
TETRA 800
iDEN 820
CDMA 850
LTE Band 5
Pulse modula
-
tion b)
18 Hz
28 V/m
28 V/m
1700 – 1990
MHz
GSM 1800
CDMA 1900
GSM 1900
DECT
LTE Band 1,2,4,25
UMTS
Pulse modula
-
tion b) 217 Hz
28 V/m
28 V/m
2400 – 2570
MHz
Bluetooth
WLAN
802.11b/g/n
RFID 2450
LTE Band 7
Pulse modula
-
tion b) 217 Hz
28 V/m
28 V/m
5100 – 5800
MHz
WLAN 802.11a/n
Pulse modula
-
tion b) 217 Hz
9 V/m
9 V/m
NOTE : If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting
antenna and the ME EQUIPMENT or ME SYSTEM may be reduced to 1m. The 1m test distance is
permitted by IEC 61000-4-3.
a) For some services, only the uplink frequencies are included.
b) The carrier shall be modulated using a 50% duty cycle square wave signal.
c) As an alternative to FM modulation, 50% pulse modulation at 18 Hz may be used because while it
does not represent actual modulation, it would be worst case.