Aqua TROLL Instruments—store within the
temperature range -40° C to +65° C (-40° F to
+149° F)

Power Options

Internal Power - Batteries

Internal batteries are not user-replaceable. The
approximate percentage of the power remaining in an
internal battery is displayed on the Home Screen when
an instrument is connected to Win-Situ

®

 Software.

External Power - External Battery Packs

External battery packs can significantly increase the life
of an instrument, either for long-term deployments or to
preserve an aging instrument.

TROLL Battery Pack (PN 0051450)

The sealed, submersible TROLL Battery Pack (lithium)
supplements internal battery power when an instrument
is used for fast, frequent sampling or during long-term
deployments. When this power source is connected, the
instrument will use the external battery source first and
switch to the internal batteries when external battery
power is depleted. Total battery life depends on the
sampling speed.

TROLL battery pack

TROLL Replaceable Battery Pack (PN
0090000)

The TROLL Replaceable Battery Pack supplements
internal battery power when a Level TROLL or Aqua
TROLL 100 or 200 Instrument is used for frequent, fast
sampling or during long-term deployments. When this
power source is connected, the TROLL instrument will
use the external battery source first and switch to the
internal batteries when external battery power is
depleted. Battery life depends on sampling rate. This
battery pack allows for 1.5 V UM-3 or size AA batteries
(8) that are replaced by the user.

TROLL Replaceable Battery Pack

Estimated Battery Lifetime

TROLL
Battery Pack

Aqua TROLL
100/200

Level TROLL
Family

15 minute
logging rate*

4.1 years

1.1 years

1 hour
logging rate*

4.4 years

1.1 years

*Logging with all sensors. Actual battery lifetime varies based on
site conditions.

TROLL
Replaceable
Battery Pack

Aqua TROLL
100/200

Level TROLL
Family

15 minute
logging rate*

5.75 years

1.6 years

1 hour
logging rate*

7 years

1.7 years

*Logging with all sensors. Actual battery lifetime varies based on
site conditions.

Antifouling Products

TROLL Shield Nose Cone (PN 0081480)

The copper TROLL Shield Nose Cone is designed to
reduce macro- and micro-fouling of the pressure sensor
on Level TROLL and Aqua TROLL Instruments.
Reduced fouling on the sensor improves measurement
accuracy and extends the length of deployments.

For optimum performance, the TROLL Shield Nose
Cone should be replaced every 12 months or sooner if
site conditions are extremely harsh.

TROLL Shield Nose Cone

800-446-7488

4

www.in-situ.com

Summary of Contents for Aqua TROLL

Page 1: ... it becomes saturated with moisture Desiccant stages from left New nearly expired replace now expired It is extremely important to use a properly sized desiccant for your deployment and to change desiccant often Desiccant should be changed before the entire volume has turned pink and you should use enough desiccant to effectively keep cables instruments and electrical boards dry until your next sc...

Page 2: ...sing the Desiccant Refill Kit 1 Remove the black nylon vent cap from the top of the desiccant 2 Pour out and discard the used desiccant Check the glass wool in the bottom of the container and replace if necessary The glass wool prevents the desiccant beads from falling out the cable end of the desiccant pack If the wool glass does this effectively there is no need to replace it when refilling the ...

Page 3: ...ch as a dilute solution of dish detergent Be careful not to damage the plastic material of the conductivity cell Instruments are shipped with polyurethane foam swabs for this purpose You can also use a thin cotton pipe cleaner and clean the cell with a gentle back and forth flossing motion Process 2 Light scrubbing with a foam swab and an aggressive soap such as Alconox detergent can be used for s...

Page 4: ...equent fast sampling or during long term deployments When this power source is connected the TROLL instrument will use the external battery source first and switch to the internal batteries when external battery power is depleted Battery life depends on sampling rate This battery pack allows for 1 5 V UM 3 or size AA batteries 8 that are replaced by the user TROLL Replaceable Battery Pack Estimate...

Page 5: ... 12 months or when the unit data appears to drift significantly Factory calibration includes a thorough cleaning full functionality check and adjustments of the pressure sensor temperature sensor and conductivity sensor when applicable over the entire calibrated temperature range To obtain a factory calibration fill out and return the online RMA Return Materials Authorization form located at www i...

Page 6: ...s in your Operator s Manual or related In Situ Inc documentation may cause damage to your instrument your warranty to become void or both Please follow all instructions and warnings to avoid damage to your product The presence of the Waste Electrical and Electronic Equipment WEEE marking on the product indicates that the device is not to be disposed via the municipal waste collection system of any...

Reviews: