VM-4424S/H Instruction Manual
(Compatible with Version 2.16 or later)
The setting method is the same as AC output sensitivity. The value is switched as shown
in Fig.4-16 by pressing the button with the number displayed.
Vel.
mm/s
Acc.
m/s
2
Disp.
mmp-p
Fig.4-16 DC Output Sensitivity Setting
Cautions when using AC output and DC output
(1)
When performing measurement using AC output and DC output, be sure to set
Auto Range to “OFF”.
If Auto Range is set to “ON”, the vibration value fluctuates greatly immediately
after when the measurement range switched. At this time, AC output and DC
output fluctuate greatly, and correct measurement results cannot be obtained.
(2) Be sure to set the sensitivity of AC output and DC output above the selected range.
Ex.
When set to the range of velocity 100mm/s,
AC Output Sensitivity is set “150”.
DC Output Sensitivity is set “150”.
4-5-2. Volume Adjustment
If you insert an earphone into the ACOUT jack, you can hear the vibration sound
during measurement.
This volume can be changed with “Volume” in Fig.4-12. It can be set in increments of
10% from 10 to 100%.
4-6. Setting of Battery Type, Auto Power Off, and Contrast
Touch “Next Page” on the setting screen shown in Fig.4-12 to move to the second page
(Fig.4-17 on the next page).
15000
1500
150
15
450
45
4.5
150
15
1.5
0.15