background image

SD Logger

  instruction manual[X519009-3] 

  21/28 

<2020/03/03> 

3

rd

 edition

 

  Other functions 

 

1.

 

Date / time setting 

 

To  set  the  date  and  time,  connect  the  logger  to  a  PC  and  execute  the  PC   

dedicated  software  (SDLG4_TADJ). 

For  details,  please  refer  to  the  separate  document 

Date  /  Time  Setting   

Instruction  Manual  <X520012>

Check  the  date  and  time  settings  after  purchase  and  before  use. 

 

2.

 

Jumper switch 

 

To set this up, remove the screw on the back of the case and remove the top cover. 
Please refer to the numbers of switches and connectors silk-printed on the board 

and the photos on the next page. 

In addition to the buzzer, this product does not come with a short-circuit jumper 

socket.

 

 
J1

Power control from outside 

By short-circuiting J1-1 or J1-2, the board power switch can be turned ON / OFF 

via the CN1 D-sub connector. 
 

Jumper shorting 

External power control terminal 

J1-1 

CN1

 

  D-Sub No.1 pin 

J1-2 

CN1

 

  D-Sub No.6 pin 

 

The board switches on at +2.5 to + 15V and switches off at -15V to + 0.8V, so it can 

be controlled at the RS-232C level or at the C-MOS level of + 3.3V to + 5V. 

To use this function, turn off the power switch (POWER / SW2). 

  D-sub pin 6 is looped back to No.4 pin on the unit. When controlling the 

power  supply  of  the  board  with  D-sub  No.6  pin,  make  sure  that  the 
handshake line of the connected device is not affected. 

 
J2

To external (or from external) power supply. 

When J2 is short-circuited, the No.9 pin of CN1 D-sub is connected to the + pin of 

CN2 DC jack, so power (8V to 12V DC) can be supplied through this pin instead of 
using the AC adapter. 

(Caution)When using the functions of J1 and J2, be sure to fully understand 
and make sure that the wiring is correct. 

 
J3

Stop the buzzer. 

Remove J3 (open) to stop the buzzer. 

 

Summary of Contents for ESD-BC4A

Page 1: ...2020 03 03 3rd edition Logger Instruction Manual ESD CS4A ESD CC4A ESD BS4A ESD BC4A Cased Case Type PCB PCB Type The contents of this instruction manual are subject to change without notice for impr...

Page 2: ...und 16 V File 17 1 File control system 17 2 Logging file script version 17 3 Subdirectory 17 4 Maximum number of files 18 5 Command version file specification Command version 18 6 Script file script v...

Page 3: ...2020 03 03 3rd edition 6 Strengthen short term power off response 25 7 Other enhancements 26 8 Main differences from Logger 3 26 IX Functions enhanced after V394X 26 1 Script log start stop function...

Page 4: ...e CMLogger provided by our company After confirming the operation on the personal computer it can be implemented in the SD logger script version Please refer to the following documents for the script...

Page 5: ...on mode In the logger the setting can be changed in Configuration Mode The items that can be changed are as follows Item Contents Function Script version Command version Data bit 1 8 Bit 7 Bit Stop bi...

Page 6: ...onnector Fixed screw Inch 4 40UNC Power supply DC 8V 12V Current consumption Standard 100mA maximum 150mA Operating temperature range 0 60 Time accuracy 0 01 Script WAIT TIME Built in clock accuracy W...

Page 7: ...art PCB type No Symbol Contents SW2 Power switch CN3 Connector for memory card LED3 Red LED Error LED2 Yellow LED Access LED1 Green LED Pilot CN2 DC Power connector DC jack CN1 RS 232C Connector D Sub...

Page 8: ...e No Symbol Contents POWER Power switch Connector for memory card ERR Red LED Error ACC Yellow LED Access PL Green LED Pirot DC8 12V DC power connector DC jack RS 232C RS 232C Connector D Sub pin male...

Page 9: ...ration setting Configuration settings are made by connecting to a PC For configuration settings refer to Chapter I 3 For details refer to the way of using ESDHost X519006 Item Dip Switch Configuration...

Page 10: ...mark Automatic clear of communication error Don t do Do Function Script version Command version Can be set Dip switch settings can be prioritized Can be set Cannot be set Default setting What has been...

Page 11: ...is set by operating dip switches 1 to 3 The correspondence between switch ON OFF and bit rate is shown in the table below No Bit rate bps 1 2 3 4 5 6 7 8 OFF OFF OFF 9 600 ON OFF OFF 19 200 OFF ON OF...

Page 12: ...Mode The mode is set by operating DIP switches 7 and 8 The correspondence between the switch ON and OFF and the mode is as shown in the table below No Mode 1 2 3 4 5 6 7 8 OFF OFF Normal ON OFF Config...

Page 13: ...o pin 6 5 GND Signal ground 6 DSR Data set ready connected to pin 4 Note 1 7 RTS Transmission request usually high level Note 3 8 CTS Transmission permission not used note 4 9 Unused Note 2 Use a cros...

Page 14: ...C power DC8V to 12V from the DC power connector DC jack CN2 for board Use a compatible DC plug with an inner diameter of 2 1 and an outer diameter of 5 5 When using an AC adapter select a suitable DC...

Page 15: ...rns off the preparation is complete Procedure 3 Power switch POWER SW2 OFF To finish the work after receiving data and exchanging commands set the power switch to the OFF side Setting the power switch...

Page 16: ...red ERR LED3 Flashes once The card is not installed Flashes twice Error in card access Flashes 3 times Data communication error Flashing In update mode The buzzer sounds as follows Buzzer When the acc...

Page 17: ...a name that does not overlap with an existing file Search for the existence of the file from 0000 LOG to FFFF LOG and the name of the first non existent file found is used as the logging file name The...

Page 18: ...e of the startup time for the number of files under certain conditions Number of files 512pcs 1 024pcs 4 096pcs 10 000pcs Start up time 1 3seconds 2 1 seconds 16 9 seconds 1minutes 29 6 seconds It is...

Page 19: ...or use with large amounts of data In the command version check the response firmly it takes longer time to get a response Notes If it takes time to write 1 byte the power will not be turned off immedi...

Page 20: ...ly is secured by the capacitor Even if the power supply main power supply to the DC power connector is cut off while the power switch is ON the power is supplied from the backup capacitor and the shut...

Page 21: ...witch can be turned ON OFF via the CN1 D sub connector Jumper shorting External power control terminal J1 1 CN1 D Sub No 1 pin J1 2 CN1 D Sub No 6 pin The board switches on at 2 5 to 15V and switches...

Page 22: ...SD Logger instruction manual X519009 3 22 28 2020 03 03 3rd edition Picture of jumper switches and power switches J2 J1 1 J1 2 J3 SW2...

Page 23: ...tes Script version Command version Receive buffer capacity 4096 4096 RTS ON threshold 768 768 RTS OFF threshold 512 512 Conflict with script function Even if flow control is enabled RTS control in the...

Page 24: ...iewing it with an editor Status file File name STATUS TXT Folder Root folder Status file format Item Example Name Version DT MC04 V400 Start up time 2018 02 15 12 57 34 Model script Communication sett...

Page 25: ...ate module file will be renamed to UPDATE OK You only need to update once and the same update will not be repeated 4 Synchronize script and incoming data In MC04 the execution of the script and the lo...

Page 26: ...een changed to be done in configuration mode FATSYS conventional FAT16 file system has been abolished and now only FatFs FAT32 compatible file system FatFs FAT32 compatible file system can also use FA...

Page 27: ...SD Logger instruction manual X519009 3 27 28 2020 03 03 3rd edition External dimensions Case type Weight about 163 PCB type Weight about 65...

Page 28: ...eriod 1 Due to incorrect use 2 Incorrect connection 3 Due to wrong power supply 4 Modified or repaired by the customer 5 Fire earthquake storm and flood damage lightning and other natural disasters po...

Reviews: