Start-up
15
5
IMT Analytics AG
The power supply socket is located at the back of FlowAnalyser. The master switch
is used to switch the device on and off. The LED, labelled Charging, is lit when the
battery is being charged. This also takes place if the device is switched off.
5.2.1
Supply voltage
The mains voltage of the power supply unit included is 100 – 240 VAC at 50 – 60 Hz.
Before switching on, make sure the operating voltage of the power
supply unit agrees with the local mains voltage.
You will find this information on the rating plate on the back of the
power supply unit.
Only operate FlowAnalyser with the original power supply unit included!
To protect the device against interference from electromagnetic fields
and static it is necessary to earth the device using the corresponding
earthing stud.
5.3.1 Filter
To protect the device against soiling due to contaminants and particles in the air the
filter included must be used for every measurement of flow (Flow High and Flow Low).
Particles of dirt in the air can clog the measuring system and thus
lead to incorrect measurements. The filter must be checked regularly
(
9.3 Preventive cleaning and servicing routines
).
5.2 Power supply
5.3 Mechanical connectors