13 | P a g e
IMR IX 176
IMR ENVIRONMENTAL EQUIPMENT, INC.
Note:
You can’t activate the detector when is being charged in the status of power off. In order to
avoid fire or explosion, please do not charge the detector when it is working in the spot.
Please try not to charge when the detector is power on, otherwise the charge speed
will be affected.
6.
DATA UPLOADING
Connect the detector to PC through the USB cable. Then open the software for uploading the
detection record.
Note:
This function can be available only when the product you
bought has the function of data communication.
7.
USING AND REPLACING OF SENSOR
As the sensor of the detector is modularized, so please pay attention to the life of the sensor.
When it is overdue, please replace it. Every half year you need to calibrate the sensor in order to
guarantee the accuracy of the detector.
The sensor replacement shoul be carried out by the seller or the local repairing department. If
there is no distributor or repairing department, please replace the sensor after you get permission
from our company. The sensor should be replaced with the same modularized sensor
supplied by our company.
8.
USING THE ACCESSORIES
For the user easy or bring, we equip buckle, crocodile buckle and flying ring with the detector. If you
need the above accessories, you can fix them on the back shell of the detector with the screw. If you
mainly use the buckle, please take out the crocodile buckle
,
and then fix the buckle.