background image

BATTERY MAINTENANCE

Install Batteries: 

Remove screw located at rear end of unit and pull down 

hinged battery door to open as shown. Always insert all four batteries into the 
battery compartment in the same direction. Note: polarity mark on the inside of the 
battery door for proper battery orientation. 

PRODUCT SPECIFICATIONS

Model #  

LD900

Name 

Leak Detector, Refrigerant Gas

Sensitivity 

.05 oz/yr R134a, .015 oz/yr R1234yf, .025 oz/yr R22

Sensor Life 

> 10 years

Response Time 

Instantaneous

Power Supply 

4 AA Alkaline batteries

Battery Life 

4-6 hours continuous

Warm up time 

< 20 seconds

Probe length 

17 inches (43.8cm)

Numerical Display 

7 segment digital display (1 to 9)

Weight, lbs. 

1.5 lbs. (0.68kg)

Warranty 

2 years (includes sensor)

EN14624-2012 TEST SPECIFICATIONS

Minimum/Maximum Sensitivity Threshold (fixed)  

1 gm/yr minimum, >50 gm/yr maximum

Minimum/Maximum Sensitivity Threshold (moving) 

3 gm/yr minimum, > 50 gm/yr maximum

Minimum Detection Time (1gm/yr) 

Approx. 1 sec

Clearing Time 

Approx. 9 seconds after exposure to >50 gm/yr

Minimum Threshold after Maximum Exposure 

1 gm/yr

Sensitivity Threshold in Polluted Atmosphere 

1 gm/yr

Calibration Frequency 

1/yr check with calibrated leak Standard

CROSS SENSITIVITY TO AUTOMOTIVE CHEMICALS

Some automotive solvents and chemicals have similar hydrocarbon properties as R134a and may elicit a positive 

response. Before leak checking, clean up any chemicals in the list below that elicit a positive response

Chemical Name/Brand 

Response

Rain-X Windshield Wash Fluid 

Y

Ford Spot Remover (Wet) 

Y

Ford Rust Inhibitor 

Y

Ford Gasket Adhesive (Wet) 

Y

Loctite Natural Blue degreaser (diluted) 

Y

Ford Brake Parts Cleaner 

Y

Ford Silicone Rubber (uncured) 

Y

Motorcraft Antifreeze heated to 160° F 

Y (partial)

Gunk liquid wrench 

Y

Ford silicone lubricant  

N

Ford Pumice lotion (with solvent) 

Y

Ford Motorcraft brake fluid  

Y

Ford Carburetor Cleaner 

Y

Dextron Transmission fluid heated to 160° F 

N

Quaker State Motor Oil heated to 160° F 

N

SENSOR MAINTENANCE

Replace Filter: 

Unscrew sensor tip as shown to replace filter. Replace filter 

whenever it becomes visibly dirty or every 2 to 3 months depending on use.

Replace Sensor: 

Remove sensor by pulling out of socket. Install the new sensor 

by aligning the Keyway notch in sensor cover with the raised keyway on sensor 
socket holder (see figure below). 

Note: 

Do not force sensor into socket. Misalignment can damage the sensor pins. 

IMPORTANT:

 

The instrument’s software is designed to alert the user if the sensor is dislodged or 

defective. If the sensor is not fully inserted into the six-pin socket, or if it is defective, the instrument will not come 
out of the “Warm Up” mode for proper operation when the power button is turned on. Additionally, if the instrument 
becomes unstable during its operation, it is an indication that the sensor may be defective or dislodged.

Unscrew tip to 

replace filter

Filter

Push straight on (do not 

twist) to install sensor

Pull straight out (do not 

twist) to remove sensor

Keyway alignment

Polarity mark

6

7

LeakDetectors-LD900-UserManual.indd   6-7

12/7/16   1:08 PM

Reviews: